Series3 Auto User Guide

05

Pump Unit  

Mattress  

Manual    

Always read this manual before using this product. 

Important safeguards 

When using electrical products, especially when 

children are present, basic safety present, basic 

safety precautions should always be followed, 

including the following:         

Danger

To reduce the risk of electrocution:
•  Always unplug this product immediately after 

use.

•  Do not use while bathing.
•  Do not place or store product where it can fall 

or be pulled into a tub or sink.

•  Do not place in or drop into water or other 

liquids.

•  Do not reach for a product that has fallen into 

water. Unplug immediately.

Warning

To reduce the risk of electrocution or injury to 

persons:

•  A product should never be left unattended 

when plugged in.

•  Close supervision is necessary when this 

product is used by, on, or near children or 

invalids.

•  Use this product only for its intended use as 

described in this manual.

•  Do not use attachments not recommended by 

the manufacturer.

• 

Never operate this product if it has a damaged 

cord or plug, if it is not working properly, if it 

has been dropped or damaged, or dropped 

into water. Return the product to a service 

centre for examination and repair.

•  Keep the cord away from heated surfaces.
• 

Never block the air openings of the product 

or place it on a soft surface, such as a bed or 

couch, where the air openings may be blocked. 

Keep the air openings free of lint, hair, and the 

like.

• 

Never drop or insert any object into any 

opening or hose.

•  Do not use outdoors or operate where 

aerosol (spray) products are being used.

•  Connect this product to a properly grounded 

outlet only. 

• 

Do not disassemble, repair or rebuild the 

product without permission, otherwise may has 

risk of fire and danger in life. 

Grounding

Before any connection is made, make certain that 

this product shall be connected to a protective 

earth conductor via the three-wire power cord; 

the three-blade plug shall be inserted only into 

a socket outlet that provides a protective earth 

contact. 

Statements

Note

Indicates some tips or some information users 

should be aware of.

Caution

Indicate correct operating or maintenance 

procedure in order to prevent damage to or 

destruction of the equipment or other property.

Introduction    

This manual should be used for the initial set up of 

the system and for reference purposes. 

General

 

The Series3 Auto is a high quality and affordable 

air mattress system support suitable for medium 

and high-risk pressure ulcer treatment. It has been 

specifically designed for prevention of bedsores 

and offers an affordable solution to 24-hour 

pressure area care.   

The Series3 Auto has been tested and certified for 

the following standards:  

•  CE
•  UL

Intended Use 

The Series3 Auto is intended to reduce the 

incidence of pressure ulcers while optimizing 

patient comfort. It also provides the following:   

•  Individual customized setting capability.

•  Pain management as prescribed by a physician.   

Contraindication 

Patient conditions for which the application of 

pressure relieving therapy on an alternation system 

is contraindicated are as follows:   

•  Cervical or skeletal traction.  
• 

Unstable spinal cord injuries. 

Package contents 

Check inside the package. If any of the following 

items should be missing or damaged, please 

contact your local dealer or service centre for 

replacement immediately.   

Version 1: 2.9.2021

User Guide 

Series3 Auto

Содержание 3 Auto Series

Страница 1: ...Version 3 17 11 2022 User Guide Series3 Auto ...

Страница 2: ...ure warning 12 Cleaning 13 Environment requirements 13 Handling and storage 13 Maintenance 15 16 Troubleshooting 17 Specifications Know your symbols Manufacturer Do not iron Foot end Tumble dry CE marked Do not dry clean REF Catalogue number Resistant to ignition REP EC Authorized EC Representative Wipe down Machine wash at 95 Recycling Drip dry UKCA marked 1 CL 1 Chlorine Medical Device Airflo Xi...

Страница 3: ...a properly grounded outlet only Do not disassemble repair or rebuild the product without permission otherwise may has risk of fire and danger in life Grounding Before any connection is made make certain that this product shall be connected to a protective earth conductor via the three wire power cord the three blade plug shall be inserted only into a socket outlet that provides a protective earth ...

Страница 4: ...on for 3 seconds to lock the panel press again for 3 seconds to unlock the panel Power Switch 1 The switch is at the right side of the pump Turn ON OFF the power and press Power button on the panel the pump will start stop operation Press Cycle Time 2 The cycle time can be selected from the panel The cycle time value options are 10 15 20 or 25 minutes Auto weight 3 Auto Press the button the pump w...

Страница 5: ... under each end of your bed Ensure buckles are securely fastened and straps are pulled tight c IMPORTANT On a profiling bed secure the side straps around the moveable sections of the bed base DO NOT secure to the side rails or main bed frame irreparable damage could be caused to the bed and mattress d Confirm there are no sharp objects in the immediate area which may risk damage to the Mattress Re...

Страница 6: ...le pressure please refer to page 8 for the hand check procedure CPR function When there is an emergency requirement to perform CPR on the patient press the CPR button at the head section of the mattress and disconnect the air hose from the pump to release the air quickly from the mattress Pressure set up The pump unit can automatically set the pressure according to patient s weight Other functions...

Страница 7: ...e to disconnect all the air cells one by one and spray the disinfectant on all sides including all the connecting tubes and hoses Let it sit for least 10 minutes If there is a base after you remove all the air cells the base has to be sprayed down with the disinfectant inside and outside Let it sit for the required incubation time and wipe down with a cloth Repeat the process with the tubing set s...

Страница 8: ...Change air cell Troubleshooting guide Problems Reasons Maintenance Mattress has pillow function but air cell fail to inflate 1 One way valve is assembled reversely 1 Assemble the one way valve in correct direction Pump is working but synchronous motor does not work thus mattress does not alternate and alternate failure alarm is activated 1 Synchronous motor is out of order 2 Wires inside synchrono...

Страница 9: ...able Nylon PVC 840D Case Material Flame Retardant ABS 20 x 8 Cell on Cell air cells Max firm mode CPR valve for emergency procedures Alarm Low pressure Power failure Alternate failure Pillow function Alternate Static Seat inflate Auto weight mode Quick connectors Size 31cm L x 12 5cm W x 21cm H Maximum Weight Capacity 200 kg Weight 2 5 kg Note The above specifications are also applicable to those ...

Страница 10: ...Series3 Auto User Guide 19 Notes Notes ...

Страница 11: ...Series3 Auto User Guide Edward House 5 Penner Road Havant PO9 1QZ 02394 318 318 ask oska uk com oska uk com Pressure Care Experts ...

Отзывы: