Series3 Auto User Guide

09

System Installation

b) Attach the air hoses through the use of the 

quick connector to the Control Unit. Ensure air 

hoses do not kink between mattress, bed frame 

and Pump Unit.

c) Plug the power cord into an electrical outlet 

with grounded AC power.

NOTE: Before inserting the plug into the outlet, 

make sure the voltage is compatible. Also make 

sure this product is well grounded.

d) Switch the power button on. Allow up to 40 

minutes for full inflation.

e) The 8 pressure indicate lights will flash until 

the mattress is inflated and ready for patient 

use. Once ready, the weight setting will show 

a constant light at the 40kg weight setting. 

Perform a “bottom out” test to ensure that 

patient is properly suspended. Slide your hand 

under the top cover along a deflated cell in 

the sacral (bottom) area. Secure sheets loosely 

enough to ensure they do not interfere with cell 

alternation.

Perform a “bottom out” test to ensure that patient 

is properly suspended. Slide your hand under 

the top cover along a deflated cell in the sacral 

(bottom) area. Secure sheets loosely enough to 

ensure they do not interfere with cell alternation. 

The following describes the procedures for setting 

up the system for the first time.

a) Remove all covers, sheets and mattress from 

the bed.

b) On a standard bed, position Mattress 

Replacement on top of bed, printed top cover 

facing upwards and air hoses towards the base 

of the bed. Attach to the bed by securing the 

adjustable straps under each end of your bed. 

Ensure buckles are securely fastened and straps 

are pulled tight. 

c) IMPORTANT: On a profiling bed, secure the 

side straps around the moveable sections of the 

bed base. DO NOT secure to the side rails or 

main bed frame – irreparable damage could be 

caused to the bed and mattress

d) Confirm there are no sharp objects in the 

immediate area which may risk damage to the 
Mattress Replacement.

Pump Unit Activation

a) Position Control Unit by hanging hooks over 

foot board of the bed or place unit on the floor 

under the bed.

Operation

NOTE: Al

ways read the operating instructions 

before use.  

General 

This product is designed to provide maximum 

comfort to patients. Make sure that you operate 

this product in a proper way optimizing its value. 

Here we provide some general information that 

you should be aware of.   

For products: 

 

• 

DO NOT use another pump with different 

specifications unless instructed to do so by 

OSKA Care      

• 

DO NOT change any component by yourself. 

If there is need for replacement or repair, 

always contact OSKA Care  

For patients: 

When the pressure has been adjusted to a desired 

level of firmness, the patient can then lie on the 

mattress.           

A firm surface will make it easier for the patient to 

transfer or reposition. Make use of the static mode 

function for this feature. To sit a patient up in bed, 

press the Seat Button to increase the pressure to 

have better support on sacrum area.

Hand check: 

Check if the pressure is properly adjusted by 

sliding one hand between the air mattress and bed 

frame to feel the patient’s buttock. 

Users should be able to feel the space in between, 

and the acceptable range is approximately 25 to 

40 mm (1” to 1-1/2”). 

Содержание 3 Auto Series

Страница 1: ...Version 3 17 11 2022 User Guide Series3 Auto ...

Страница 2: ...ure warning 12 Cleaning 13 Environment requirements 13 Handling and storage 13 Maintenance 15 16 Troubleshooting 17 Specifications Know your symbols Manufacturer Do not iron Foot end Tumble dry CE marked Do not dry clean REF Catalogue number Resistant to ignition REP EC Authorized EC Representative Wipe down Machine wash at 95 Recycling Drip dry UKCA marked 1 CL 1 Chlorine Medical Device Airflo Xi...

Страница 3: ...a properly grounded outlet only Do not disassemble repair or rebuild the product without permission otherwise may has risk of fire and danger in life Grounding Before any connection is made make certain that this product shall be connected to a protective earth conductor via the three wire power cord the three blade plug shall be inserted only into a socket outlet that provides a protective earth ...

Страница 4: ...on for 3 seconds to lock the panel press again for 3 seconds to unlock the panel Power Switch 1 The switch is at the right side of the pump Turn ON OFF the power and press Power button on the panel the pump will start stop operation Press Cycle Time 2 The cycle time can be selected from the panel The cycle time value options are 10 15 20 or 25 minutes Auto weight 3 Auto Press the button the pump w...

Страница 5: ... under each end of your bed Ensure buckles are securely fastened and straps are pulled tight c IMPORTANT On a profiling bed secure the side straps around the moveable sections of the bed base DO NOT secure to the side rails or main bed frame irreparable damage could be caused to the bed and mattress d Confirm there are no sharp objects in the immediate area which may risk damage to the Mattress Re...

Страница 6: ...le pressure please refer to page 8 for the hand check procedure CPR function When there is an emergency requirement to perform CPR on the patient press the CPR button at the head section of the mattress and disconnect the air hose from the pump to release the air quickly from the mattress Pressure set up The pump unit can automatically set the pressure according to patient s weight Other functions...

Страница 7: ...e to disconnect all the air cells one by one and spray the disinfectant on all sides including all the connecting tubes and hoses Let it sit for least 10 minutes If there is a base after you remove all the air cells the base has to be sprayed down with the disinfectant inside and outside Let it sit for the required incubation time and wipe down with a cloth Repeat the process with the tubing set s...

Страница 8: ...Change air cell Troubleshooting guide Problems Reasons Maintenance Mattress has pillow function but air cell fail to inflate 1 One way valve is assembled reversely 1 Assemble the one way valve in correct direction Pump is working but synchronous motor does not work thus mattress does not alternate and alternate failure alarm is activated 1 Synchronous motor is out of order 2 Wires inside synchrono...

Страница 9: ...able Nylon PVC 840D Case Material Flame Retardant ABS 20 x 8 Cell on Cell air cells Max firm mode CPR valve for emergency procedures Alarm Low pressure Power failure Alternate failure Pillow function Alternate Static Seat inflate Auto weight mode Quick connectors Size 31cm L x 12 5cm W x 21cm H Maximum Weight Capacity 200 kg Weight 2 5 kg Note The above specifications are also applicable to those ...

Страница 10: ...Series3 Auto User Guide 19 Notes Notes ...

Страница 11: ...Series3 Auto User Guide Edward House 5 Penner Road Havant PO9 1QZ 02394 318 318 ask oska uk com oska uk com Pressure Care Experts ...

Отзывы: