Series3 Auto User Guide
09
System Installation
b) Attach the air hoses through the use of the
quick connector to the Control Unit. Ensure air
hoses do not kink between mattress, bed frame
and Pump Unit.
c) Plug the power cord into an electrical outlet
with grounded AC power.
NOTE: Before inserting the plug into the outlet,
make sure the voltage is compatible. Also make
sure this product is well grounded.
d) Switch the power button on. Allow up to 40
minutes for full inflation.
e) The 8 pressure indicate lights will flash until
the mattress is inflated and ready for patient
use. Once ready, the weight setting will show
a constant light at the 40kg weight setting.
Perform a “bottom out” test to ensure that
patient is properly suspended. Slide your hand
under the top cover along a deflated cell in
the sacral (bottom) area. Secure sheets loosely
enough to ensure they do not interfere with cell
alternation.
Perform a “bottom out” test to ensure that patient
is properly suspended. Slide your hand under
the top cover along a deflated cell in the sacral
(bottom) area. Secure sheets loosely enough to
ensure they do not interfere with cell alternation.
The following describes the procedures for setting
up the system for the first time.
a) Remove all covers, sheets and mattress from
the bed.
b) On a standard bed, position Mattress
Replacement on top of bed, printed top cover
facing upwards and air hoses towards the base
of the bed. Attach to the bed by securing the
adjustable straps under each end of your bed.
Ensure buckles are securely fastened and straps
are pulled tight.
c) IMPORTANT: On a profiling bed, secure the
side straps around the moveable sections of the
bed base. DO NOT secure to the side rails or
main bed frame – irreparable damage could be
caused to the bed and mattress
d) Confirm there are no sharp objects in the
immediate area which may risk damage to the
Mattress Replacement.
Pump Unit Activation
a) Position Control Unit by hanging hooks over
foot board of the bed or place unit on the floor
under the bed.
Operation
NOTE: Al
ways read the operating instructions
before use.
General
This product is designed to provide maximum
comfort to patients. Make sure that you operate
this product in a proper way optimizing its value.
Here we provide some general information that
you should be aware of.
For products:
•
DO NOT use another pump with different
specifications unless instructed to do so by
OSKA Care
•
DO NOT change any component by yourself.
If there is need for replacement or repair,
always contact OSKA Care
For patients:
When the pressure has been adjusted to a desired
level of firmness, the patient can then lie on the
mattress.
A firm surface will make it easier for the patient to
transfer or reposition. Make use of the static mode
function for this feature. To sit a patient up in bed,
press the Seat Button to increase the pressure to
have better support on sacrum area.
Hand check:
Check if the pressure is properly adjusted by
sliding one hand between the air mattress and bed
frame to feel the patient’s buttock.
Users should be able to feel the space in between,
and the acceptable range is approximately 25 to
40 mm (1” to 1-1/2”).