Series3 Auto User Guide
13
Cleaning
In this section, we describe the procedures to clean
and decontaminate the pump. It is important to
follow these procedures before using the system
again.
The cleaning task is required at least once a week
to maintain personal hygiene.
Pump Unit
•
DO NOT immerse or soak the pump in any
water or fluids.
•
Check for external damage and move the
pump to the cleaning area.
• Place the pump on a work surface and
spray or wipe the outside of the case with
quaternary ammonium solution.
•
DO NOT spray any cleaning solution directly
on the surface of the pump.
•
DO NOT use a Hypocarbonate or Phenolic
based cleaning solution as this may cause
damage to the case. Allow the solution to
incubate for 10 minutes or accordingly as
stated by the cleaning product instructions.
• Wipe case with a clean cloth. Make sure all
areas are clean (top and bottom, both sides).
• Spray cloth with cleaning solution and clean
faceplate. DO NOT allow excess cleaning
solution on faceplate or control panel. (If
solution gets inside, damage will occur.) Allow
surface to thoroughly dry after cleaning.
• After the pump is thoroughly cleaned and
dried, proceed to plug in the pump and test to
see if it runs normally.
• Unplug the pump and store with proper
identification tag.
•
Avoid long exposure to sunlight.
Mattress
• Brush off or wipe down all surfaces of the
cover sheet with soap and water before
wetting with any liquid disinfectant.
• Any obvious blood spots should be wet
thoroughly with 1:9 Hypochiorite solution (1
part bleach to 9 parts water) and allow drying
for at least 10 minutes. Then blot with a clean,
damp cloth.
• Unzip the top cover from the mattress.
• Brush or wipe down all surfaces with soap and
water before applying any liquid.
• Covers are immersed and soaked in
disinfectant for the required incubation time.
•
After pre-soaking, the cover is rinsed through
a regular cycle in a washer with no soap
then laundered with mild detergent (wash
temperature 93°F/34°C, rinse temperature
78°F/26°C or on the coldest setting).
• Covers are aerated until they are fully dry.
(Drying temperature range 90-120°F/32-49°C
or on the coldest setting.)
• The air cells are unsnapped from one side and
are sprayed on all sides with a disinfectant.
Let it sit for the required incubation time and
wipe down with a clean cloth. (Make sure to
disconnect all the air cells, one by one, and
spray the disinfectant on all sides, including all
the connecting tubes and hoses. Let it sit for
least 10 minutes. )
• If there is a base after you remove all the air
cells, the base has to be sprayed down with
the disinfectant, inside and outside. Let it sit for
the required incubation time and wipe down
with a cloth.
•
Repeat the process with the tubing set: spray,
incubate, and then wipe clean.
• The carrying bag should be turned inside
out and completely wiped down using the
disinfectant solution. Allow it to thoroughly air
dry. Once the inside is dry, turn it back: wipe
down the outside of the bag with disinfectant.
•
Dry the mattress on a flat surface area after
cleaning, away from exposure to the sun.
•
Avoid long exposure to sunlight.
Environment requirements
Operating Conditions
•
Ambient Temperature: 5°C~40°C
•
Relative Humidity: 15%~90%, non-condensing
•
Atmospheric pressure: 700hPa to 1060hPa
Storage and shipping conditions
•
Ambient temperature: -25°C~70°C
•
Relative Humidity: 10%~90%, non-condensing
•
Atmospheric pressure: 700hPa to 1060hPa
Handling and storage
•
Lay the mattress out flat and upside down.
• Roll from the head end towards the foot end;
the foot-end strap can then be stretched
around the rolled mattress to prevent unrolling.
•
Do not fold, crease or stack the mattress.
Maintenance
General
• Check the power cord and plug to see if there
are abrasions or excessive wear.
• Check the mattress cover for signs of wear or
damage. Ensure the mattress cover and tubes
are connected together correctly.
•
Plug in the pump unit and check the airflow
from the hose connection port. The airflow
should alternate between ports every half-
cycle time.
• Check the air hoses to see if there are any
kinks or breaks. For replacement, please
contact your local agent or dealer.
• Make sure the mattress tube is well connected.
• Check the pump unit and make sure both
power indicators are off when the switch is
turned off.
Low pressure
Examine if there is any air leakage between the
pump and the mattress connections or from the
air mattress tubes:
• Check connectors between the air mattress
and pump. If there is any disconnection, please
reconnect it.
• Check the CPR Valves. Ensure their outlets are
sealed.
• Check the air-connecting tubes. Ensure each
single cell is not broken.
•
Set the pressure at Max firm. Keep the tubes
fully inflated and inspect for air leakage.
• Check if there is any air leakage from cells.
Ensure no leakage occurs. If any leakage occurs,
please contact your local agent or dealer.