![OrthoAccel Technologies AcceleDent Aura Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/orthoaccel-technologies/acceledent-aura/acceledent-aura_directions-for-use-manual_1654591026.webp)
26
Direction for Use AcceleDent® Aura
Recommended Separation Distance
The Activator is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the
user of the Activator can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Activator as
recommended below, according to the maximum output power of the
communications equipment.
150 KHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
0.01
0.1
1
10
100
0.12
0.38
12
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.