4. FIRMWARE COMMANDS AND RESPONSES
33
SHOW_INTEGRAL_PRESET
Reports the current ROI integral preset value. For more information about the ROI integral
preset, see SET_INTEGRAL_PRESET. See also SHOW_INTEGRAL.
Responses:
$G0000000000075<CR>
Integral preset reported as 0.
...
...
$G4294967295132<CR>
Integral reported as 4294967295.
SHOW_LIVE
Reports the contents of the live-time counter in units of 20 ms (50 ticks/s). See also
CLEAR_COUNTERS and SET_LIVE.
Responses:
$G0000000000075<CR>
Live time reported as 0 ticks.
$G0000000001076<CR>
Live time reported as 1 tick (20 ms).
...
...
$G4294967295132<CR>
Live time reported as 4294967295 ticks (over 23000
days).
SHOW_LIVE_PRESET
Reports the current live-time preset in units of 20 ms (50 ticks/s). See also
CLEAR_PRESETS and SET_LIVE_PRESET.
Responses:
$G0000000000075<CR>
Live-time preset reported as disabled.
$G0000000001076<CR>
Live-time preset reported as 1 tick.
...
...
$G4294967295132<CR>
Live-time preset reported as 4294967295 ticks.
SHOW_LLD
Shows the lower level discriminator setting. See also SET_LLD.
Responses:
$C00050092
The LLD is 50.
SHOW_MDA [PeakNumber]
Reports current MDA for the MDA Preset selected by PeakNumber. PeakNumber is
assumed to be 0 if not supplied.
Example response:
MDA 0000000000010.7
SHOW_MARK
Not used by the EASY-MCA; included for backward compatibility.