2. GETTING STARTED
11
Figure 9. EASY-MCA About Tab.
Figure 10. EASY-MCA Presets Tab.
2.3.2. About
This tab (Fig. 9) displays hardware
and firmware information about the
currently selected EASY-MCA, as
well as the data Acquisition Start
Time and Sample description. In
addition, the Access field shows
whether the MCB is currently pass-
word locked (see the password dis-
cussion in the MAESTRO user
manual). Read/ Write indicates
the MCB is unlocked; Read Only
means it is locked.
2.3.3. Status
The Status tab is not used in the
EASY-MCA.
2.3.4. Presets
Figure 10 shows the Presets tab.
MDA presets are on a separate tab.
The presets can only be set on an
MCB that is not acquiring data
(during acquisition the preset fields
are disabled). You may use any or
all of the presets at one time. To
disable a preset, enter a value of
zero. If you disable all of the pre-
sets, data acquisition will continue
until manually stopped.
The values of all presets for the cur-
rently selected MCB are shown on
the Status Sidebar. These values do not change as new values are entered on the Presets tab; the
changes take place only when you Close the Properties dialog.
When more than one preset is enabled (set to a non-zero value), the first condition met during
the acquisition causes the MCB to stop. This can be useful when you are analyzing samples of