5.  CALIBRATION AND MAINTENANCE

5.1.  TEST MODE

If the jumper link marked W3 on the PWB is removed,
the system will run in a test mode in which the timers
will run at 60 times the normal rate. For example, a
period set to 3 hours will actually elapse in 3 minutes.
This is useful for checking system operation.

WARNING Remember to reinstall W3 after

testing.

5.2.  STORAGE AND SHIPPING

DANGER

Isolate the controller from the mains
power supply before removing the
cover.

With the CryoSecure disconnected from the mains
supply, remove the top cover and locate the sealed,
rechargeable, lead-acid battery (to the left of the unit as
viewed from the front). Disconnect the red lead to the
battery, wait 10 seconds, then re-connect it. The battery
will now be isolated from the control circuitry until mains
power is reapplied.

Содержание CryoSecure

Страница 1: ...CryoSecure Compressor Power Controller Hardware Manual Printed in U S A ORTEC Part No 774890 1209 Manual Revision D...

Страница 2: ...o that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the...

Страница 3: ...THE POWER ON SAFE STATE 4 3 3 INSTALLATION IN A WARM SYSTEM 4 3 4 INSTALLATION IN A COLD SYSTEM 4 4 OPERATING INSTRUCTIONS 4 4 1 OPERATIONAL STATES 4 4 1 1 Normal Operation 4 4 1 2 AC Mains Power Fai...

Страница 4: ...zard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please read al...

Страница 5: ...v...

Страница 6: ...vi...

Страница 7: ...output to the compressor is limited to 600 W No power conditioning or conversion is performed The CryoSecure incorporates a microprocessor controlled circuit to detect a mains power failure measure t...

Страница 8: ...the detector bias voltage to be applied DANGER Isolate the controller from the mains power supply before removing the cover DEEP DISCHARGE PROTECTION In the event of a long term power failure or if th...

Страница 9: ...normal working state bypassing the warming and cooling states by pressing and holding the INITIALIZE button for 5 seconds when in the cooling bias off state 4 OPERATING INSTRUCTIONS 4 1 OPERATIONAL ST...

Страница 10: ...e the CryoSecure waits after a mains power failure before restarting the compressor is set with the two rotary selector switches mounted along the top edge of the controller PWB labeled WARM HI and WA...

Страница 11: ...on WARNING Remember to reinstall W3 after testing 5 2 STORAGE AND SHIPPING DANGER Isolate the controller from the mains power supply before removing the cover With the CryoSecure disconnected from the...

Страница 12: ...6...

Отзывы: