background image

4

4.1.4.  The Cooling State
In the cooling state, the AC POWER and BIAS OFF
LEDs are lit while the MAINS FAIL, HOLDING, and
WARMING LEDs are off. The compressor power is
reapplied, but the detector bias supply is held off.

The system will remain in the cooling state for the
period set on the internal BIAS OFF TIME switch, after
which the BIAS OFF LED will go out and the detector
bias voltage can be reapplied. The system will then
return to  normal operation (see Section 4.1.1).

The cooling period (bias-off time) is set to a value long
enough for the detector to be cooled to a temperature
where it is safe to re-apply the bias voltage.

If the mains supply fails during the cooling period, the
CryoSecure will return to the start of the warming state.

4.2.  SETTING THE HOLDING TIME

The holding time, which is the time the CryoSecure
waits after a mains failure before going into the warming
state, is set with the two rotary selector switches
mounted along the top edge of the controller PWB,
labeled HOLD HI and HOLD LO.

DANGER

Isolate the controller from the mains
power supply before removing the
cover.

The switches are BCD-type selectors, each with
settings from 0 to 9. The number set on the switches
corresponds to the holding time, with each unit of the
number set representing 5 minutes. The following table
give examples of some possible settings.

HOLD HI

HOLD LO

Time

(min)

Time

(h:min)

0

1

5

0:05

1

0

50

0:50

3

5

175

2:55

9

9

495

8:15

4.3.  SETTING THE WARMING TIME

The warming time, which is the time the CryoSecure
waits after a mains power failure before restarting the
compressor, is set with the two rotary selector switches
mounted along the top edge of the controller PWB,
labeled WARM HI and WARM LO.

The switches are BCD-type selectors, each with
settings from 0 to 9. The number set on the switches
corresponds to the warming time, with each unit of the
number set representing 1 hour. The following table
gives examples of some possible settings:

WARM HI

WARM LO

Time (h)

0

1

1

1

0

10

3

5

35

9

9

99

4.4.  SETTING THE COOLING

(BIAS-OFF) TIME

The cooling, or bias-off, time is the time the CryoSecure
waits after restarting the compressor before allowing the
detector bias voltage to be reapplied. It is set with the
rotary selector switch mounted along the top edge of the
controller PWB, labeled BIAS OFF.

The switch is a hexadecimal-type selector with settings
from 0 to F. The number set on the switches
corresponds to the holding time, with each unit of the
number set representing 2 hours. The following table
give a list of the possible settings:

Switch Setting Time (h) Switch Setting

Time (h)

0

0

8

16

1

2

9

18

2

4

A

20

3

6

B

22

4

8

C

24

5

10

D

26

6

12

E

28

7

14

F

30

4.5.  DEEP DISCHARGE PROTECTION

In the event of a long-term power failure or if the mains
is switched off for a long period, the battery, normally
charged to 6.75 V, will eventually discharge to 5.25 V.
When this voltage is reached, a self-holding relay will
de-energize, disconnecting the battery. The CryoSecure
will then be in the shipping/storage-preparation state
described in Section 5.2. The battery will reconnect
automatically when mains power is restored, and the
system will go into the “safe” state described in
Section 3.2.

Содержание CryoSecure

Страница 1: ...CryoSecure Compressor Power Controller Hardware Manual Printed in U S A ORTEC Part No 774890 1209 Manual Revision D...

Страница 2: ...o that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the...

Страница 3: ...THE POWER ON SAFE STATE 4 3 3 INSTALLATION IN A WARM SYSTEM 4 3 4 INSTALLATION IN A COLD SYSTEM 4 4 OPERATING INSTRUCTIONS 4 4 1 OPERATIONAL STATES 4 4 1 1 Normal Operation 4 4 1 2 AC Mains Power Fai...

Страница 4: ...zard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please read al...

Страница 5: ...v...

Страница 6: ...vi...

Страница 7: ...output to the compressor is limited to 600 W No power conditioning or conversion is performed The CryoSecure incorporates a microprocessor controlled circuit to detect a mains power failure measure t...

Страница 8: ...the detector bias voltage to be applied DANGER Isolate the controller from the mains power supply before removing the cover DEEP DISCHARGE PROTECTION In the event of a long term power failure or if th...

Страница 9: ...normal working state bypassing the warming and cooling states by pressing and holding the INITIALIZE button for 5 seconds when in the cooling bias off state 4 OPERATING INSTRUCTIONS 4 1 OPERATIONAL ST...

Страница 10: ...e the CryoSecure waits after a mains power failure before restarting the compressor is set with the two rotary selector switches mounted along the top edge of the controller PWB labeled WARM HI and WA...

Страница 11: ...on WARNING Remember to reinstall W3 after testing 5 2 STORAGE AND SHIPPING DANGER Isolate the controller from the mains power supply before removing the cover With the CryoSecure disconnected from the...

Страница 12: ...6...

Отзывы: