background image

3

3.  INSTALLATION

3.1.  CONNECTION TO PROPORTIONAL

COUNTER

A direct connection with shielded cable should be
made between the proportional counter and the
Input SHV connector on the preamplifier. The
performance of the 142PC Preamplifier, like that of
all other such low-noise preamplifiers, is degraded
as the capacity at the input increases. Therefore it
is important that the length of coaxial cable used
between the proportional counter and the
preamplifier be kept at the minimum length that is
necessary. Also it is preferable to use 93

S

 or 100

S

characteristic impedance cable rather than 75

S

 or

50

S

 cable because the capacity per foot is less for

the cable with the higher characteristic impedance.
Type RG-62/U cable is recommended; this has a
93

S

 impedance and a capacitance of 13.5 pF/ft

(40.1 pF/m). An AMP 51426-2 connector mates
with the SHV connector on the 142PC Preamplifier.

Once the input cable installation has been made,
the electronic noise performance of the 142PC can
be predicted by calculating the cable capacity from
the above information and adding the capacity
expected from the proportional counter. 

3.2.  CONNECTION TO AMPLIFIERS

 
Either or both the E and T outputs of the 142PC can
be connected to an amplifier input for further
processing. The output impedance through either of
these connectors is 93

S

 providing series

termination for 93

S

 cable so that long unterminated

cable lengths can be driven easily. Although the
outputs are marked E (for Energy) and T (for
Timing), the pulse characteristics are identical and
the circuits are interchangeable. 

In an energy spectrometer system, the preamplifier
output is furnished into a shaping main amplifier. In
a timing spectrometer system, the preamplifier
output is furnished into a timing filter amplifier. With
the dual output connections on the 142PC, the
signals can be furnished simultaneously to both
types of spectrometer systems. If either the E or T
output connector is not being used, it should simply
be left open-circuited (unterminated). 

3.3.  INPUT POWER

 
Power for the 142PC is supplied through the Power
Cable that is captive through the rear panel of the
unit. The normal connection for this power cable is
included on the rear panel of the mating ORTEC
amplifier, furnishing ±12 V and ±24 V from the bin
and power supply in which the amplifier is
operating. If this facility is not available or if such a
connection would increase the loading on the bin
and power supply beyond its maximum rated
capacity, use an ORTEC 114 Preamplifier Power
Supply to furnish the operating power requirements
through the captive cable. The ORTEC 114 can
furnish power for two preamplifiers simultaneously
if desired. 

3.4.  TEST PULSE

 
A voltage test pulse for energy calibration can be
accepted through the Test Pulse input connector on
the 142PC without the use of an external charge
terminator. The test input of the preamplifier has an
input impedance of 93

S

 and its circuitry provides

charge injection to the preamplifier input. The
shape of this pulse should be a fast risetime (less
than 40 ns) followed by a slow exponential decay
back to the baseline, (200 to 400 

:

s). While the test

pulses are being furnished, connect either the
detector (with bias applied) or the equivalent
capacitance (without bias applied) to the Input
connector on the 142PC. 

The Test Pulse input may be used for the output of
a pulser such as the ORTEC 419 or 448 to calibrate
the preamplifier E output amplitude in terms of
energy for multichannel analyzer calibration.
However, due to stray coupling between the test
circuit and other portions of the preamplifier
circuitry, the transient performance of the
preamplifier is best determined by connecting the
actual proportional counter signal through the Input
connector instead of using the pulse generator
output signals for this calibration. 

A voltage test pulse for transient response in the
142PC can be accepted through a charge
terminator and into the Input connector on the
142PC. If external capacitance is to be included for

Содержание 142PC

Страница 1: ...Model 142PC Preamplifier Operating and Service Manual Printed in U S A ORTEC Part No 717600 1202 Manual Revision B...

Страница 2: ...n so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of t...

Страница 3: ...2 2 5 ELECTRICAL AND MECHANICAL 2 3 INSTALLATION 3 3 1 CONNECTION TO PROPORTIONAL COUNTER 3 3 2 CONNECTION TO AMPLIFIERS 3 3 3 INPUT POWER 3 3 4 TEST PULSE 3 3 5 DETECTOR BIAS INPUT 4 4 OPERATION 4 4...

Страница 4: ...iv...

Страница 5: ...e Input connector on the preamplifier 2 Discharge the bias circuitry before making any connections to the Input connector and before disconnecting the preamplifier from the proportional counter 3 To d...

Страница 6: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 7: ...the instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 8: ...viii...

Страница 9: ...y long cable lengths can be used between the proportional counter and the preamplifier without disrupting the system stability An internal risetime compensation adjustment is accessible through a hole...

Страница 10: ...rtional counter BIAS Accepts the bias voltage for the proportional counter from a bias supply TEST PULSE Accepts input voltage pulses from a pulse generator for instrument and system check and calibra...

Страница 11: ...ther the E or T output connector is not being used it should simply be left open circuited unterminated 3 3 INPUT POWER Power for the 142PC is supplied through the Power Cable that is captive through...

Страница 12: ...should be furnished with the bias supply module 4 OPERATION 4 1 GENERAL When the 142PC is installed according to the appropriate information in Section 3 it operates at all times when power is applie...

Страница 13: ...The energy output signal from the preamplifier is a fast risetime voltage step with an exponential return to the baseline in about 75 s The polarity of signals through the E output is inverted from t...

Страница 14: ...timing experiments typical risetimes of 25 to 150 ns for detector capacities from 1 to 100 pF the total cable length should be kept as short as possible even though it is terminated Due to vagaries i...

Страница 15: ...ue only to the noise contribution of the preamplifier and main amplifier The analyzer can be calibrated in terms of keV per channel by observing two different pulser peaks of known energy and the FWHM...

Страница 16: ...hed charge termination to measure the rise time of the 142PC through the T or E output connector Connect the 419 output through the charge terminator to the input of the 142PC and use an oscilloscope...

Отзывы: