General Information 

 
This manual is intended for: Transporters - Installers - Users - Maintenance and Disposal personnel 
Please read this manual carefully and in its entirety, as it contains important information relating to the safe and efficient operation of the device. 
This manual is an integral part of the product and is included only for the device to which it relates. 
This manual should be stored securely for the entire useful life of the device so that it may be referred to as needed. In the event a used device 
is sold, it must be sold together with the manual and any appendices. The manufacturer assumes no liability for direct or indirect damages to 
persons, property or animals caused by use of the device in conditions other than those set out in this manual. The manufacturer reserves the 
right to make changes without notice to this documentary material or to the device to which it refers. 
 
Safety pictograms: 

 

Danger indicates a condition or situation that may cause death or serious injury. 

 

Warning indicates a condition or situation that may cause moderate injuries

.

 

Note indicates additional information, explanations or hints.

 

 

Identification 

Manufacturer:  

 

OROTIG Srl 

 

Via XXV Aprile 47 

 

37014 

 Cavalcaselle di Castelnuovo del Garda 

 

Ph.: +39 045 6400865, FAX +39 045 6401104 

 

E-mail: [email protected] 

 web: www.orotig.com 

 
Device: 

L0F102HZ  

LASER WELDER MOD. LDT REVO-S, SPOT 0,15 C/W. CAMERA. 

L0F102HN  

LASER WELDER MOD. LDT REVO-S PLUS, SPOT 0,2 C/W. CAMERA. 

L0F200HZ  

LASER WELDER MOD. REVO 15000-S, SPOT 0,15 C/W. CAMERA 

L0F200HN  

LASER WELDER LDT 15000 REVO-S PLUS, SPOT 0,2 C/W. CAMERA 

 
The identification plate on the device displays its electrical characteristics and complete identification information. 
 
 

 

Содержание Revo S L0F102HN

Страница 1: ...ury to the operator if used correctly and properly However no engineering design can make this device safe if it is not used and maintained properly and in compliance with safety standards This manual...

Страница 2: ...ort 10 Unpacking 10 Positioning 10 Preliminary checks 10 Electrical connections 11 Power supply tolerances 11 Device specifications 12 Technical specifications 12 Applicable standards and directives 1...

Страница 3: ...the protective glass slide 22 Routine Maintenance 22 Removal of the liquid 22 Filling refilling the liquid 22 Filling up inserting liquid 22 Checking Cleaning of rear filters 23 Check resin filters 2...

Страница 4: ...itions other than those set out in this manual The manufacturer reserves the right to make changes without notice to this documentary material or to the device to which it refers Safety pictograms Dan...

Страница 5: ...ure to laser radiation can cause cataracts and in the worst cases burning of the retina Do not remove the hand curtains for any reason Persons of small stature should not be near the hand curtains whe...

Страница 6: ...lding chamber Bottles and objects containing gas or highly flammable liquids should be moved a safe distance away from the welding chamber Keep a fire extinguisher nearby in case of fire Do not use wa...

Страница 7: ...m the welding chamber and to prevent the operator or people close by from exposure to the flashes produced during welding operations as well as to prevent any leakage of shielding gas in use Operation...

Страница 8: ...ng some device operation procedures These risks can be eliminated by following basic work and behaviour rules or more specifically the procedures in this manual and by using personal protective equipm...

Страница 9: ...very and provide an explanation The customer should be careful when using box cutters or other cutting tools when opening the package Package content The package contains the following items 1 Laser W...

Страница 10: ...oning The device should be positioned in a location and environment suited to the use for which it was designed laboratory use sheltered from the weather such positioning should be performed by qualif...

Страница 11: ...led and sized according to the specifications of the device Power supply tolerances Voltage at 10 of the rated voltage Frequency 1 of the rated voltage for continuous use 2 of the rated voltage for sh...

Страница 12: ...hour Tested at 2kW 2ms 10Hz with room temperatures of 25 C 77 F Continuous Dimensions max Stereo microscope Leica with 10X eyepieces Leica with 10X eyepieces The device offers the ability to adjust th...

Страница 13: ...fhydrocarbon gas anymixtureof Hydrogenand Nitrogen Do not use Nitrogen when welding Do not place any flammable materials along the path of the laser beam Do not use any toxic materials or materials th...

Страница 14: ...ct the energy level of the laser generator used in this device is Class 4 5 Enabling Key Safety devices on equipment are adopted LASER radiation protection glass KG3 under microscope Leather stripes f...

Страница 15: ...ctions see section 6 Rear connections see section 7 Liquid cooling racks do not cover the grill leave at least 5 cm of space between the grid and the wall 8 Coolant level window 9 Cooling circuit filt...

Страница 16: ...no defective damaged or worn parts perform any regular maintenance or repairs that may be required Positioning and ergonomics Microscope eyepieces Remove the caps that protect the stereo microscope i...

Страница 17: ...e welding area through the coaxial chamber of the stereo microscope in this mode a crosshair is displayed on the screen to indicate the point of impact of the laser beam Switching between the display...

Страница 18: ...t controls to adjust the desired height then set the desired parameters on the display according to the metal to be welded Use the control pedal to perform the welding Emergency Stop The emergency but...

Страница 19: ...menu is protected by password and is accessible only to personnel authorized for the repair of the device 7 Cam This button switches the view to the welding chamber after a few seconds of inactivity o...

Страница 20: ...r touching anywhere on the touch screen the display returns to the welding parameters 13 Location of the memory cell pages there are 70 memory cells displayed 7 at a time Refer to chapter Memory Cells...

Страница 21: ...cone in the cone holder to direct the flow to the welding point Be sure that the lever 1 is turned to a position that releases the gas from the conveyor cone 2 It is recommended to use compressed air...

Страница 22: ...lutions Routine Maintenance The routine maintenance operations described in this chapter do not required special training they can be performed by any operator provided that the directions concerning...

Страница 23: ...chine removing the 6 screws shown in the image Unscrew the caps blocking the filters and slide them out Replace them with new ones code RIC00254 Cleaning the filters Failure to maintain these filters...

Страница 24: ...erials Leave sufficient space at least 10cm between the aspiration grills and walls or obstacles so as not to compromise the efficiency of the machine Replacing the protective glass slide To access th...

Страница 25: ...itate the alignment procedure the crosshairs bear the symbol which indicates the highest point make sure this symbol is pointed upwards and complete the alignment as described above Only use the viewi...

Страница 26: ...ndition of side filters and clean or replace if necessary If automatic recovery does not happen contact technical support WARNING Cod 6 CCPS Hot 1 Problem The temperature of the electronic circuits fo...

Страница 27: ...open and not blocked Check the condition of side filters and clean or replace if necessary If automatic recovery does not happen contact technical support ERROR Cod 16 Vcap High 2 Problem The voltage...