45 Wrong product combination:
Mixup between non compatible device and hand held
programming unit
->
Use the correct hand held programming unit (Pos. 14)
46 Handset error internal communication:
Invalid interchip communication inside the hand held
programming unit
->
Replace the hand held programming unit (Pos. 14)
47 Internal communication error motor control:
Internal communication error motor control.
->
Replace motor control (Pos. 20)
48 User stoped the special function
49 Unknown error in the motor control:
->
Replace motor control (Pos. 20)
3.2 Mechanics
The movable screws should not be completely unscre-
wed when adjustments are being made. Make sure
that the movable screws are tightened for operation
and transport.
The frame is unstable:
Possible cause: Bolt / screws missing or loose. Tighten
the screws / bolts.
3.3 Others
Do not clean the housing or the support with grease or
oil.
No solvents may be used when cleaning the
OptiFlex-K1.