10. Checklist of safety and function test
OptiFlex-K1
Safety test
Protective earth conductor resistance
Ground leakkage current EN 60601 / IEC 601/ VDE 0751
Or
Ground leakkage current as in UL 2601
Function test
≤
≤
≤
Measured value
Date/ Signature
0,1 Ohm
Ohm
500 µA
µA
300 µA
µA
OK
Error
1.
Switch on the OptiFlex -K1. Press the two outer control
(Extension/ Flexion)
simultaneously.
Display: Software version VX.X XX.XX.XX (X = optional) Keep on pressing.
Display: Artromot K1 Classic
2. The maximum range of motion for Extension/ Flexion is -10 to 120 degrees.
Check the angle in position 0 degrees. Tol/- 5 degrees.
Check the angle in position 60 degrees. Tol/- 5 degrees.
Check the angle in position 100 degrees. Tol/- 5 degrees.
3. Check the emergency-off function.
Start the OptiFlex K1 in any mode.
Press any control, the OptiFlex- K1 will stop immediately.
Check this for all controls.
5. Start the OptiFlex -K1 in the motion range between –10 to 120 degrees.
Set the speed to 100%.
Both extreme points should be reached within 45 – 65 seconds.
13
4. Checking the "PAUSE" function.
Select the following settings:
Extension = 10 degrees
Flexion
= 90 degrees
Speed
= 50%
Pause
= 10 seconds
Start the OptiFlex-K1.
At the reversal points (at 10 / 90 degrees) the pause must be lasting 10 seconds
(+/- 2 seconds).
The speed is substantially lower than in a 100% setting.