Thank you very much for your trust in an Actius by Orliman 

product. You have purchased a quality product. Please read 

the instructions carefully before using the product and use 

the product only as described and for the intended purposes.
If you have any questions about the use, maintenance or sa-

fety of your product, contact the healthcare establishment 

where you purchased the product or our customer service 

department. Keep these instructions for future reference. 

Read these instructions before using the product.

REGULATIONS

This product is compliant with the European Directive on Me-

dical Devices 93/42/EEC (Spanish Royal Decree 1591/2009). A 

Risk Analysis (EN ISO 14971) was done, minimising all exis-

ting risks. 
Strictly observe municipal ordinance when discarding the pac-

kaging and product.

SUITABLE FOR

This mattress is recommended to prevent pressures ulcers up 

to Grade I (ACDE01#UNI), Grade II (ACDE02#UNI) and as an 

auxiliary method to treat patients who spend long periods of 

time in bed.  It is used as a mattress cover; therefore a base 

mattress is required for proper use. The user must not weigh 

more than the maximum weight indicated.

COUNTERINDICATIONS

Patient conditions for which therapy using an alternating 

pressure system is not recommended: 

· Cervical or skeletal traction. 

· Unstable marrow dysfunction.

PRODUCT DESCRIPTION:
Pump

1. Switch: The switch is on the right side of the pump. By 

turning the switch ON/OFF, the Start/Stop pump begins/

stops operating. 2. Air valves: Connect the air tubes to the 

air valves. 3. Pressure adjustment button: Turn the Min./Max. 

button to a comfortable pressure level.

Characteristics 

• Output ACDE01 5-8 L/min, ACDE02 6-7 L/min 

• Adjustable pressure button. 

• 6-minute cycle. 

• Built-in bed assembly supports. 

• 2m power cable.

Alternating pressure mattress 

• Dimensions: 

200cm Long x 90cm Wide x 7cm High 

• 130 individual bubble cells for max. surface area 

• Straps to position and secure the alternating pressure 

mattress. 

*• Maximum user weight: ACDE01 120 kg, ACDE02 145 kg

USE

Step 1. Position the alternating pressure mattress. The air 

tubes must be at the foot of the bed where the air valves on 

the left side of the pump can be easily connected.
Step 2. Cover it with a cotton sheet to prevent direct contact 

with the skin and reduce friction.
Step 3. Hang the pump over the frame at the foot of the bed. 

Make sure the pump is secure.
Step 4. Connect the alternating pressure mattress air 

tubes to the pump air valves. Make sure they are correctly 

connected.

ATTENTION: Make sure the air tubes are not bent or under 

the alternating pressure mattress.
Step 5. Plug in the power cable
NOTE: Before plugging in, make sure the voltage is compa-

tible.
Step 6. Connect the power by pressing the on/off switch to 

the right of the pump. See the operating section.
Step 7. Make sure the pump is turned off and the power 

cable is unplugged when not in use.

STEP-BY-STEP OPERATION:

Pump APP
Step 1. Turn it on. The switch light will turn on. The pump will 

begin pumping air into the pillow.

Step 2. The first time it is used, turn the pressure button all 

the way up.
Step 3. After fully inflating it, turn the pressure button and 

adjust the pressure from gentle to strong in accordance with 

the patient’s weight and preferences.
Set the pressure. The pressure button should be at the maxi-

mum setting when the pillow is blown up for the first time. 

Users can then easily adjust the pillow pressure level to the 

desired firmness or as recommended by a professional.

If the pressure is low, check for any leaks in the air tubes. 

Contact your local distributor to replace damaged tubes.

CLEANING:

Pump
Do not submerge or soak the pump in water. Check for any 

external damage and move the pump to a cleaning area. 

Place the pump on a work surface and clean the outside of 

the box with a quaternary ammonium solution. Do not use 

any cleaning product directly on the pump surface. Do not 

use any hypocarbonate or phenol-based cleaning product as 

it could damage the box.

The solution should remain on the product for 10 min. or as 

indicated on the cleaning product used. Clean the front with 

a cloth. Do NOT allow any excess cleaning product on the 

control panel or front of the box. If the product gets inside 

the box, it will damage the pump. Let the surface completely 

dry after cleaning.
When the pump is completely clean and dry, plug in the 

pump and test it to guarantee normal operation.
Alternating pressure mattress
Any blood stain must be moistened with a 1:9 hypochlorite 

solution (1 part bleach and 9 parts water).  Let dry for at 

least 10 min. Then dry with a clean, damp cloth. Let air dry 

completely.
Once dry inside, turn it over: clean the outside of the alterna-

ting pressure mattress.

PRECAUTIONS

The construction material is inflammable. Do not expose the 

products to situations that could set them on fire. In the event 

of a fire, quickly let go of them and use the proper resources 

to put it out. Do not expose the product to direct heat sources 

such as stoves, heaters, radiators, etc.

RECOMMENDATIONS AND WARNINGS

Follow these safety measures when electrical products are 

used and particularly in the presence of children: 

DANGER 

 (to reduce the risk of electrical shock).

· Unplug immediately after use. 

· Do not use while bathing. 

· Do not place or store the product where it may fall into 

water. 

· Do not place or let fall in water or any other liquid. 

· Do not touch the product after it has fallen into water. 

Immediately unplug.

WARNINGS 

 (to reduce the risk of burns, electrical 

shock, fire or injury).

· The product may not be left alone when plugged in. 

· The product must be supervised by an adult and kept out of 

the reach of children. 

· Use this product only for the intended use as described in 

this manual. Do not use accessories not recommended by 

the manufacturer. 

· A broken cable or socket.  If it does not work properly or has 

fallen, been broken or submerged in water, call an authorised 

distributor to repair or replace it. 

· Keep the power cable far from hot surfaces. 

· Do not cover air outlets on the product or leave on any type 

of soft surface such as a bed or sofa where the air outlets 

may get blocked.  

· Keep the air outlets free of fuzz, hair and dust. 

· Do not let it fall or insert any object into any opening or the 

hose. 

· Do not use outside or where aerosols (sprays) or oxygen 

are used. 

· Connect this product only to earthed sockets. See the 

instructions.

STORAGE AND MAINTENANCE

STORAGE

 

Smooth out the alternating pressure mattress. Carefully fold 

or roll. Unplug the pump and store with the identification la-

bel. Follow national regulations when disposing of the pump.

MAINTENANCE

 

Periodically check the power cable and socket for any 

abrasions or excessive wear. Plug in the pump and check the 

air flow from the hose connection port. The air flow must 

alternate between ports every half-cycle.

Keep in a dry place at room temperature.

WARRANTY

ORLIMAN, S.L.U. guarantees all its products as long as the 

original configuration has not been manipulated or altered. It 

does not guarantee any products with altered characteristics 

due to improper use, defects or breakage of any kind. The 

compressor comes with a 2-year warranty and the mattress, 

a 6-month warranty, for material or manufacturing defects. 

Excluded from the warranty are imperfections caused by im-

proper use, accidents or natural wear of parts such as plastic 

components, footrests, upholstery, brakes, tips, inner tubes 

and wheelchair tyres and similar parts, depending on the type 

of medical product acquired.

If you do not follow the instructions, warnings and the notes 

in the user’s manual, you or a third party may be physically 

injured or the product may be damaged. 
If you observe any defect, anomaly or failure, notify the dea-

ler immediately so it may be replaced or repaired.

ANTI-BEDSORE MATTRESS 
Instructions for use and care

ORLIMAN S. L. U. 

C/ Ausias March, 3 - Pol. Ind. La Pobla-L´Eliana
Apdo. de correos 49 · C.P.: 46185
La Pobla de Vallbona Valencia - España (Spain)

MADE IN SPAIN

EDICIÓN: 2019-03

LBIACDE#1 / LBIACDE#2

ACDE01#UNI

ACDE02#UNI

Length

2000 mm

2000 mm

Height

70 mm

115 mm

Width

900 mm

900 mm

Maximum occupant weight

120 kg

145 kg

Dear Customer,

ACDE01#UNI

ACDE02#UNI

ACDE01#UNI

ACDE02#UNI