ORDISI ASPIT Скачать руководство пользователя страница 7

 

Pág.

 

7 de 9

 

 

Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11 

<5%

U

T  

 (>95% dip in 

U

T) 

 for 

0,5 cycle 
<40%

U

T  

 (>60% dip in 

U

T) 

  

for 5 cycles 
<70%

U

T  

 (>30% dip in 

U

T) 

 

for 25 cycles 
<5%

U

T  

 (>95% cycles 

U

T) 

  

for 5 s 

Ídem  

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the 
equipment requires continued operation 
during power mains interruptions, it is 
recommended that the equipment be 
powered from an uninterruptible power 
supply or a battery. 

Power frequency (50/60 Hz) 
magnetic field IEC  61000-4-

3 A/m 

3 A/m 

The power frequency magnetic field 
should be measured in the intended 
installation location to assure 

 
 

 

VACCUM VARIATION  IN RELATION TO ALTITUDE 

ABOVE SEA LEVEL, AT A 

TEMPERATURE

 

OF

 

22ºC

 

Altitude above sea 

level 

ASPIT Model 

Ph (mmhg) 

Ph (kPa) 

0 m 

630 

80 

500 m 

566 

75,5 

1000 m 

534 

71,2 

1500 m 

504 

67,2 

2000 m 

476 

63,4 

2500 m 

449 

59,8 

3000 m 

424 

56,5 

Содержание ASPIT

Страница 1: ...are i ntre inere Maintenance and Functioning manual Mom 09 Edici n 12 Fecha edici n 2019 11 04 Tecnologia 2 Nave 7 08780 Palleja Barcelona Spain 34 93 334 01 12 FAX 34 93 440 25 64 comercial ordisi co...

Страница 2: ...or directly with ORDISI S A ORDISI S A Tecnologia 2 08780 Palleja Barcelona Espa a Tel 34 93 334 01 12 Fax 34 93 440 25 64 e mail comercial ordisi com 2 MANUFACTURER S LIABILITIES O OR RD DI IS SI I...

Страница 3: ...rflowing and penetration of liquid in the pump interior will thus be prevented In this case we advise you to remove the pump lathe head for to clean it by an Authorized Technical Service 2 Excessively...

Страница 4: ...has the same indications The variation is that the inlet diameter of the plug is higher 8 5 mm 5 CLEANING AND STERILIZING 5 1 BODY Before undertaking any operation disconnect the equipment from electr...

Страница 5: ...rements 340x150x300 Weight 4 600 Kg Power 105 VA Voltage 230V 50 Hz Fuses 2 of 2A type F Consult us for other voltage Equipment of I Class B type Average life 10 years 8 SPECIAL ACCESSORIES ON OFF ped...

Страница 6: ...re not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The equipment is suitable for use in all establishments including domestic establishments not indus...

Страница 7: ...continued operation during power mains interruptions it is recommended that the equipment be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000...

Страница 8: ...P g 8 de 9 ASPIT...

Страница 9: ...P g 9 de 9 ASPIT G...

Отзывы: