10.- TROUBLESHOOTING
FAILURE OR MALFUNCTION
CAUSE
SOLUTION
Equipment starts up but doesn’t
aspirate.
Regulation control is open.
Close control.
Suction tube wrongly inserted..
Insert correctly.
Cap improperly adjusted.
Adjust properly.
Equipment starts up but suction is not
enough or inexistent.
Bacterial filter is saturated.
Change filter.
Failure continues.
Entry of liquid inside the pump.
Disassembly and cleaning of pump
head by Technical Services.
Equipment doesn’t start up nor the
luminous switch lights up.
E
xamine the fuse possible deterioration and
the power plug.
Replace the fuse or the plug.
Failure continues.
?
Contact Technical Services.
11.- TECHNICAL SPECIFICATION, EMC INFORMATION
Guidance and manufacturer’s declaration – electromagnetic emissions
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions CISPR 11
Group 1
The equipment uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The equipment is suitable for use in all establishments, including
domestic establishments (not industrially) and those directly connected
to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
I
MMUNITY
test
IEC 60601
test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power supply lines
±1KV for input/output lines
±2kV for power supply
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.