4.- INSTALLATION AND COMMISSIONING 

 

4.1.- UNPACKING 

 
Upon reception of the equipment, if package is seen as damaged, immediately check if device has been affected. If 
damages are confirmed, immediately refer them to sender. 
 

4.2.- COMMISIONING 

 

Not to use in explosive atmospheres like the anesthetic produced ones by inflammable. 
 
1)  Check room electrical installation where equipment is tobe fitted it such installation meets 

IEC (UNE)

 standards 

with a good earthing. Check voltage is same as show in the nameplate 
 

2)  Press ON/OFF light-switch Nº 1. Light-switch should now be lighted and device start its operation. 

 

3)  With  the  aspirator,  you  will  find  a  aspiration  set  for  a  single  use,  (

GRAVE  RISK  OF  INFECTION  IF  USED  IN 

MANY  PATIENTS) 

composed  by  2  meters  of  suction  tube,  one  probe,  and  connector.  The  union  of  the  probe 

and the suction tube, is through the grooving connector,  by pressure. Connect the other extreme of the tube, in 
the connector patient, as shown in the plug of the bottle Nº 3, and turn on the aspirator. To see point 2 section 
4.3. 
References for new orders 

US0110020

 3 meters suction tube (20 units)  

US0214100

 aspiration probe CH14 (100 units)  

US1000020

 conical connector (20 units) 

FB-194A      

Bacterial filter 

 

4)  Through vacuum regulator control Nº 8, is possible to increase or to decrease vacuum level. 

 

5)  Vacuum  gauge  Nº  2  will  show,  at  any  time,  the  vacuum  level  being  actually  caused  on  the 

patient.(tolerance±1.6%).  It  is  recommended  that  in  normal  conditions,  the  calibration  of  the  vacuum-gauge  is 
made once every two years. 

 

4.3.- RECOMENDATIONS FOR USE  

 

1)  Avoid  an excessive  filling  of the collection  bottle  (recommended  not  to  fill  more  that  80% of  his capacity), 

a  possible  overflowing  and  penetration  of  liquid  in  the  pump  interior  will  thus  be  prevented.  In  this  case,  we 
advise you to remove the pump lathe head for to clean it, by an Authorized Technical Service. 
 

2)  Excessively  large  suction  tubes  should  never  be  used,  because  they  have  a  trend  to  nipping  and    any 

subsequent  collapse  will cause  a suction   choking. On the other hand, a risk is run that elbows are made up   
wherein aspirated   fluids  can  be  settled  originating   therefore a  possible   infection  and  pollution centre. 
Periodically replace the tube of aspiration by a specific tube for this one assignment (it must take to noticeable 
CE  of  conformity  fulfilling  the  norm  UNE  EN  1617  of  drainage  tubes).  Verify  when  replacing  these  and  before 
coming to the  
We recommended the use of the following tubes of aspiration:   
 
-  PVC tube minimum inner diameter 6 mm by 9 mm of outer diameter.                              
-  Silicone tube minimum inner diameter 6 mm by 10 mm of outer diameter. 
 

3)  Avoid patient resting on the suction tubes in a way that suction might be choked. 

 

4)  Clean and sterilize the top Nº 5, the safety valve Nº 6 and the bottle, after of each use. 

 

5)  Never run the aspirator if a vacuum is detected in the circuit, previously have vacuum level decreased  to 0. 

 

6)  This equipment have not been designed for   continuous drainage. Do not  use it for this purpose. 

 

7)  Is advisable to use an  antifoaming for  to avoid  that the accumulation of foaming in the bottle, covers the filling 

level. 

 

8)  Never  connect  any mechanism  nor arrangement  (as ventilators, aerosol  generators, etc.) to  the air  outlet. 

 

9)

  In case of  exce

ssive filling of bottle and  if take the safety valve Nº 6, the aspirator don’t work. To stop the 

equipment with the switch, decreased  the vacuum level to 0 with the vacuum regulator Nº 8 and to change or to 
empty the bottle.

 

 

10)

 

The 

biological risk waste will be removed following the protocol of the hospital waste management

 

Содержание ASPIT

Страница 1: ...are i ntre inere Maintenance and Functioning manual Mom 09 Edici n 12 Fecha edici n 2019 11 04 Tecnologia 2 Nave 7 08780 Palleja Barcelona Spain 34 93 334 01 12 FAX 34 93 440 25 64 comercial ordisi co...

Страница 2: ...or directly with ORDISI S A ORDISI S A Tecnologia 2 08780 Palleja Barcelona Espa a Tel 34 93 334 01 12 Fax 34 93 440 25 64 e mail comercial ordisi com 2 MANUFACTURER S LIABILITIES O OR RD DI IS SI I...

Страница 3: ...rflowing and penetration of liquid in the pump interior will thus be prevented In this case we advise you to remove the pump lathe head for to clean it by an Authorized Technical Service 2 Excessively...

Страница 4: ...has the same indications The variation is that the inlet diameter of the plug is higher 8 5 mm 5 CLEANING AND STERILIZING 5 1 BODY Before undertaking any operation disconnect the equipment from electr...

Страница 5: ...rements 340x150x300 Weight 4 600 Kg Power 105 VA Voltage 230V 50 Hz Fuses 2 of 2A type F Consult us for other voltage Equipment of I Class B type Average life 10 years 8 SPECIAL ACCESSORIES ON OFF ped...

Страница 6: ...re not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The equipment is suitable for use in all establishments including domestic establishments not indus...

Страница 7: ...continued operation during power mains interruptions it is recommended that the equipment be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000...

Страница 8: ...P g 8 de 9 ASPIT...

Страница 9: ...P g 9 de 9 ASPIT G...

Отзывы: