ORDISI ASPIT Скачать руководство пользователя страница 5

 

Pág.

 

5 de 9

 

 

 
6.- ENTERTAIMENT AND MAINTENANCE 

 

 
6.1.- BACTERIA FILTER CHANGE 

 
The  bacterial  filter  included  in  your  equipment  is  a  closed  system  filter,  It  is  NON  REUSABLE, 

(GRAVE  RISK  OF 

INFECTION IF USED IN MANY PATIENTS)

.  

 

NEVER USE YOUR ASPIRATOR WITHOUT BACTERIA FILTER 

 

6.2.- FUSES CHANGE 

 

If device is connected to the power supply and switch Nº 1 is at position “1”, there is not any operation and the switch 
is not lighted, check then the fuses condition following up these instructions: 
 
1)  Remove plugs of fuseholders  Nº 10 with a screw-driver. 

 

2)  If damaged, replace fuses with news ones (Ref. RC009954) 

 

3)  Replace plugs at original position.  

 
 

7.- TECHNICAL CHARACTERISTICS 

 

 

 

 

 USING SYMBOLS 

 

 

High vacuum aspirator. 

 

Low flow aspirator. 

 

Pump of double membrane. 

 

Pump flow: 800 l/h. 

 

Max. vacuum: 84 Kpa (630 mmHg). 

 

Glass bottle capacity: 1’6 litres. 

 

Measurements: 340x150x300 

 

Weight: 4,600 Kg.  

 

Power: 105 VA  

 

Voltage: 230V / 50 Hz  

 

Fuses: 2 of 2A, type F. 

 

(Consult us for other voltage). 

 

Equipment of I Class B type. 

 

Average life: 10 years 

 

 

8.- SPECIAL ACCESSORIES 
 

 

ON/OFF pedal , with electrical microswitch , and 3,5 meters of electrical cable; Ref. A017078. 

 

Hydrophobic bacteria filter; Ref. FB-193, bag 10 units. 

 
 

9.- DIAGRAMS AND DRAWINGS 
 

At  the  request  of  the  customer,  ORDISI,  S.A.  will  be  supply  components  lists,  description  of  its,  calibration 
instructions and whatever necessary information for to repair the equipment by qualified technical personnel. 
In the present manual, we herewith enclose the electrical diagram and one complete exploded view of the equipment. 

 

9.1.- PNEUMATIC DIAGRAM 

 

1.- Patient. 
2.- Collecting bottle. 
3.- Safety valve 
4.- Bacteria filter 
5.- Vacuum regulator 
6.- Vacuum gauge 
7.- Motor-pump  
8.- Air outlet

  

 
 
 

Consult accompaniment documents. 

 

Protection earth (inside). 

 

Equipment of B class. 

 

Year of manufacture 

 

CE registered for notified organization 
 
Selective collection of electrical apparatuses 
 

 

 

Содержание ASPIT

Страница 1: ...are i ntre inere Maintenance and Functioning manual Mom 09 Edici n 12 Fecha edici n 2019 11 04 Tecnologia 2 Nave 7 08780 Palleja Barcelona Spain 34 93 334 01 12 FAX 34 93 440 25 64 comercial ordisi co...

Страница 2: ...or directly with ORDISI S A ORDISI S A Tecnologia 2 08780 Palleja Barcelona Espa a Tel 34 93 334 01 12 Fax 34 93 440 25 64 e mail comercial ordisi com 2 MANUFACTURER S LIABILITIES O OR RD DI IS SI I...

Страница 3: ...rflowing and penetration of liquid in the pump interior will thus be prevented In this case we advise you to remove the pump lathe head for to clean it by an Authorized Technical Service 2 Excessively...

Страница 4: ...has the same indications The variation is that the inlet diameter of the plug is higher 8 5 mm 5 CLEANING AND STERILIZING 5 1 BODY Before undertaking any operation disconnect the equipment from electr...

Страница 5: ...rements 340x150x300 Weight 4 600 Kg Power 105 VA Voltage 230V 50 Hz Fuses 2 of 2A type F Consult us for other voltage Equipment of I Class B type Average life 10 years 8 SPECIAL ACCESSORIES ON OFF ped...