The instruments communicate via serial line RS232 or RS485. For communication they use the ASCII protocol. Communica-
tion runs in the following format:
ASCII:
8 bit, no parity, one stop bit
The transfer rate is adjustable in the instrument menu and depends on the control processor used. The instrument address is
set in the instrument menu in the range of 0 ÷ 31. The manufacture setting always presets the ASCII protocol, rate of 9600
Baud, address 00. The type of line used - RS232 / RS485 - is determined by an exchangeable card automatically identified
by the instrument.
COMMANDS FOR INSTRUMENT OPERATION
The commands are described in specification you can find at
www.orbit.merret.cz/rs
.
A command consists of a number and a letter. The size of the letters have a significance.
DETAILED DESCRIPTION OF COMMUNICATION VIA SERIAL LINE
Activity
Data transferred
Data solicitation (PC)
#
A
A
<CR>
Data transfer (Instrument)
>
R
<SP>
D
D
D
D
D
(D)
(D)
<CR>
Command corfi rmation (Instrument) - OK
!
A
A
<CR>
Command corfi rmation (Instrument) - Bad
?
A
A
<CR>
Instrument identifi cation
#
A
A
1Y
<CR>
HW identifi cation
#
A
A
1Z
<CR>
One-time mesasurement
#
A
A
7X
<CR>
Repeated mesasurement
#
A
A
8X
<CR>
Setting to transmit d relay value
#
A
A
1X
<CR>
Setting to transmit measured value
#
A
A
1x
<CR>
Setting limit1
#
A
A
1L
D
(D)
(D)
(D)
(D)
(D)
<CR>
Setting limit 2
#
A
A
2L
D
(D)
(D)
(D)
(D)
(D)
<CR>
LEGENDA
#
35
23
H
Beginning of the command
A
A
0…31
Two signs of the inst. address
(sending in ASCII - decades and
units, ex. “01“, “99“ universal
<CR>
13
0D
H
Carriage return
<SP>
32
20
H
Space
D
Data - usually signs “0“…“9“, “-“,
“.“; (D) - DP and (-) may prolong
data
R
50
H
…57
H
Relay and Tare status
!
33
21
H
Positive command corfi rmation (ok)
?
63
3F
H
Negative command corfi rmation
(bad)
>
62
3E
H
Beginning of the transmitted data
8
DATA PROTOCOL
RELAY, TARE
Signs
Relay 1
Relay 2
Tare
P
0
0
0
Q
1
0
0
R
0
1
0
S
1
1
0
T
0
0
1
U
1
0
1
V
0
1
1
W
1
1
1
Содержание OM 652UC
Страница 1: ...OM 652UC 6 DIGIT PROGRAMMABLE IMPULSE COUNTER FREQUENCY METER STOPWATCH TIMER GUARANTEE Y E A R S ...
Страница 21: ...INSTRUCTIONS FOR USE OM 652UC 21 5 SETTING light light ...
Страница 27: ...INSTRUCTIONS FOR USE OM 652UC 27 6 SETTING profi profi ...
Страница 39: ...INSTRUCTIONS FOR USE OM 652UC 39 6 SETTING profi profi ...
Страница 43: ...INSTRUCTIONS FOR USE OM 652UC 43 6 SETTING profi profi ...
Страница 49: ...INSTRUCTIONS FOR USE OM 652UC 49 6 SETTING profi profi ...
Страница 55: ...INSTRUCTIONS FOR USE OM 652UC 55 8 DATA PROTOCOL ...
Страница 57: ...INSTRUCTIONS FOR USE OM 652UC 57 ...
Страница 59: ...INSTRUCTIONS FOR USE OM 652UC 59 10 TECHNICAL DATA ...
Страница 62: ...62 INSTRUCTIONS FOR USE OM 652UC NOTE ...
Страница 63: ...INSTRUCTIONS FOR USE OM 652UC 63 NOTE ...