Electrical Installation
53
-
The industrial process interface provides the following inputs and outputs:
Name
Description
max range
1)
/ status
A IN 1 / 2
Analog input 1 and 2
0-10 V
2)
D IN 1
Digital input
(active-low = 0…0,6 V)
24 V
AO1 / 2 / 3
Analog output 1, 2 and 3
Alarm output 1, 2 and 3
0/4-20 mA
DO1 / 2/ 3
Relay output 1, 2 and 3
3)
open/ closed (red LED on) / 0...30 V, 400 mA
FS
Fail-safe relay
open/ closed (green LED on)/ 0...30 V, 400 mA
1)
depending on supply voltage; for 0-20 mA on the AO the PIF has to be powered with min. 5V < (1.5 + working resistance * 0.021) <
24 V; Example: R
Load
= 500 ohm
→ U
min
= 1.5 + 500 * 0.021 = 12 V, R
Load
= 100 ohm
→ U
min
= 1.5 + 100 * 0.021 = 3.6 V
→ min. 5 V
2)
the AI is designed for max. 24 V, the voltage level above 10 V is not interpreted
3)
active if AO1, 2 or 3 is/ are programmed as alarm output
The alarm output can be configured as a threshold between
0-4 mA
for
no alarm
and between
10-20 mA
as
alarm
. For values outside the respective range, the relay does not switch on the
DO.
Содержание PI 640
Страница 1: ...Operator s Manual optris PI 400i 450i 450i G7 640 640 G7 05M 08M 1M Infrared camera ...
Страница 29: ...Mechanical Installation 29 Figure 5 PI 400i PI 450i PI 450i G7 dimensions mm ...
Страница 30: ... 30 Figure 6 PI 640 PI 640 G7 optics 29 33 53 60 dimensions mm ...
Страница 31: ...Mechanical Installation 31 Figure 7 PI 640 PI 640 G7 optics 13 15 dimensions mm ...
Страница 32: ... 32 Figure 8 PI 640 PI 640 G7 optics 80 90 dimensions mm ...
Страница 33: ...Mechanical Installation 33 Figure 9 PI 640 microscope optics 10 12 dimensions mm ...
Страница 34: ... 34 Figure 10 PI 05M PI 08M PI 1M dimensions mm ...
Страница 37: ...Mechanical Installation 37 Figure 16 Focusing screws for focus ring ...
Страница 40: ... 40 Figure 19 CoolingJacket Dimensions ...
Страница 42: ... 42 Standard Version Figure 22 CoolingJacket Advanced Part No ACPIxxxCJAS Standard Version Dimensions ...
Страница 44: ... 44 Figure 25 CoolingJacket Advanced Part No ACPIxxxCJAE Extended Version Dimensions ...
Страница 58: ... 58 Figure 36 Stand Alone operation with PI Netbox Figure 37 USB Server Gigabit ...
Страница 64: ... 64 6 2 Software window Figure 39 Software window 1 2 2 2 3 4 5 6 7 8 9 10 11 ...
Страница 73: ...Basics of Infrared Thermometry 73 Figure 43 Non contact thermometry ...
Страница 80: ... 80 Appendix A Table of emissivity for metals ...
Страница 81: ...Appendix A Table of emissivity for metals 81 ...
Страница 82: ... 82 Appendix B Table of emissivity for non metals ...
Страница 90: ... 90 Figure 52 Relay output at industrial PIF ...
Страница 91: ...Appendix G Declaration of Conformity 91 Appendix G Declaration of Conformity ...
Страница 92: ...optris PI MA E2020 05 A ...