48
-
4.1 Process interface
The process interface (electronics within cable as well as industrial interface) must be powered
separately (5-24 VDC). Before switching on the power the PIF cable must be connected to the
camera.
The PI is equipped with a process interface (cable with integrated
electronics and terminal block), which can be programmed via the
software as an Analog Input (AI) and Digital Input (DI) in order to
control the camera or as an Analog Output (AO) in order to control
the process. The signal level is always 0-10 V (DI = 24 V).
The process interface can be activated choosing the following options:
Analog Input (AI):
Emissivity, ambient temperature, reference temperature, uncommitted value, flag control, triggered recording,
triggered snapshots, triggered linescanner, triggered event grabber, reset peak-/value-hold, switch temperature
range
Analog Output (AO):
Main measure area, measure area, internal temperature, flag status, recording status, line scan status, alarm,
frame sync, fail-safe, external communication
Digital Input (DI):
Flag control, triggered snapshots, triggered recording, triggered linescanner, triggered event grabber, reset peak-
/value-hold, switch temperature range
Содержание PI 640
Страница 1: ...Operator s Manual optris PI 400i 450i 450i G7 640 640 G7 05M 08M 1M Infrared camera ...
Страница 29: ...Mechanical Installation 29 Figure 5 PI 400i PI 450i PI 450i G7 dimensions mm ...
Страница 30: ... 30 Figure 6 PI 640 PI 640 G7 optics 29 33 53 60 dimensions mm ...
Страница 31: ...Mechanical Installation 31 Figure 7 PI 640 PI 640 G7 optics 13 15 dimensions mm ...
Страница 32: ... 32 Figure 8 PI 640 PI 640 G7 optics 80 90 dimensions mm ...
Страница 33: ...Mechanical Installation 33 Figure 9 PI 640 microscope optics 10 12 dimensions mm ...
Страница 34: ... 34 Figure 10 PI 05M PI 08M PI 1M dimensions mm ...
Страница 37: ...Mechanical Installation 37 Figure 16 Focusing screws for focus ring ...
Страница 40: ... 40 Figure 19 CoolingJacket Dimensions ...
Страница 42: ... 42 Standard Version Figure 22 CoolingJacket Advanced Part No ACPIxxxCJAS Standard Version Dimensions ...
Страница 44: ... 44 Figure 25 CoolingJacket Advanced Part No ACPIxxxCJAE Extended Version Dimensions ...
Страница 58: ... 58 Figure 36 Stand Alone operation with PI Netbox Figure 37 USB Server Gigabit ...
Страница 64: ... 64 6 2 Software window Figure 39 Software window 1 2 2 2 3 4 5 6 7 8 9 10 11 ...
Страница 73: ...Basics of Infrared Thermometry 73 Figure 43 Non contact thermometry ...
Страница 80: ... 80 Appendix A Table of emissivity for metals ...
Страница 81: ...Appendix A Table of emissivity for metals 81 ...
Страница 82: ... 82 Appendix B Table of emissivity for non metals ...
Страница 90: ... 90 Figure 52 Relay output at industrial PIF ...
Страница 91: ...Appendix G Declaration of Conformity 91 Appendix G Declaration of Conformity ...
Страница 92: ...optris PI MA E2020 05 A ...