Air Purge Collars [LT/ 2WLT15/ 2WLT15H/ 2W2M]
The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid
reading errors. These effects can be reduced by using an air purge collar. Make sure to use oil-free,
technically clean air, only.
A combination of the
Laminar air purge collar
with the bottom section of
the
Mounting fork
allows
an adjustment in two axes.
[ACCTMG]
Standard air purge collar;
Laminar air purge collar – the sideward
fits to the mounting bracket;
air outlet prevents a cooling down of the object
hose connection: 3x5 mm
in short distances; hose connection: 3x5 mm
[ACCSAP]/ for heads with a
[ACCTAPL]
D:S ratio
≥
10:1
The needed amount of air (approx. 2...10 l/ min.) depends on the application
and the installation conditions on-site.