30
-
5 Electrical Installation
5.1 Cable Connections
Basic version
The basic version is supplied without connection cable. To connect the
CSlaser please open at first the sensor backplane (4 screws). Please use a
4-wire shielded cable which you have to conduct through the cable gland.
During assembling please make sure the shield gets a safe electrical
contact to the sensor housing. For an easier connection the terminal block
can be removed from the PCB by pulling off.
Shield connection to cable gland