14
-
2ML
2MH
G5HF
Temperature range (scalable)
250...800 °C
385...1600 °C
200…1650 °C
Spectral range
1,6 µm
5,0 µm
Optical resolution
150:1
300:1
45:1
System accuracy
1)
±(0,3 % of r2 °C)
3)
±1 °C or ±1 %
2)
Repeatability
1)
±(0,1 % of r1 °C)
3)
±0,5 °C or ±0,5 %
2)
Temperature resolution (NETD)
0,1 K
2)
Response time (90% signal)
10 ms
30 ms
Warm-up time
-
10 min
Emissivity/ Gain
0,100...1,100 (adjustable via switches on sensor or via software)
IR window correction
0,100...1,100 (adjustable via software)
Signal processing
Average, peak hold, valley hold, extended hold functions with threshold and
hysteresis (adjustable via software)
1)
at ambient temperature 23
5 °C; whichever is greater
2)
at object temperatures >0 °C
3)
= 1/ Response time 1 s