background image

is to take the tooth-like part of the nosepiece to turn it, and make the objective 

into correct position and into the bright path.   

 

3.2.6    Use of tension adjustment knob    (Fig.4) 

    The  tension  adjustment  ring  can  adjust  the  tension  of  the  coarse  and  fine 

focusing  unit  to  prevent  the  stage  from  sliding  down  automatically  and  to 

improve the comfort of operation. Rotation clockwise makes tension decrease, 

and by contraries, rotation counterclockwise makes tension increase. 

3.2.7    Use of oil immersion objective 

The 100X objective can be used for observation even if without immersion oil. 

But  little  non-synthetic  resin  immersion  oil  or  distilled  water  between  the 

objective  and  specimens  will  fully  grasp  the  objective  function  ,  bubbles  and 

impurities can’t in immersion oil which will affect image. 

after  adjusting  with  the  40X  objective,  adding  moderate  immersion  oil 

between the front  lens of 100X objective and the cover glass of specimen can 

make the image clearer. Please pay attention that air bubble and impurity can’t 

be in the immersion oil, otherwise, the image would be affected. 

▲After immersion oil used, the oil of specimen and the microscope surface 

should be cleaned with absorbent cotton, lens paper, gauze or soft cotton cloth 

with moderate mixture of pure industrial alcohol and ether (proportion 1:4) 

▲Standard thickness 0.17mm cover glass should be chosen when high times 

objective used, and thickness error should be within 0.01mm, otherwise, image 

definition wouldbe affected. 

 

Содержание A11.2601

Страница 1: ...A11 2601 Student Biological Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Страница 2: ...1 CONTENTS Attentions 2 1 Application 3 2 Specifications 4 3 Installation and use 4 4 Maintenance 10 5 Troubleshooting 13...

Страница 3: ...asoline paper and other flammable materials can t near the lamp in case of fire 4 Notes on replacing the bulb Replacement should be based on the identity of the instrument using the same specification...

Страница 4: ...tyle steady structure convenient operation and clear image is suitable for observing various biological specimens and mostly applied in school college hospital etc Binocular head Nosepiece Objective C...

Страница 5: ...Nosepiece Quadruple Condenser Abbe Stage Mechanical stage Fixed stage Optional Focusing Unit Fine adjusting unit division 0 002mm Illumination Mirror Optional 3W LED 12V20W Halogen Optional Accessory...

Страница 6: ...ter with switch used by the instrument is for protecting lamp in case high electric current shock prolong service time 3 2 2 Place sample Fig 2 Please place specimen on the stage and the face with cov...

Страница 7: ...ield coincide completely Interpupillar distance is different for everyone so interpupillar distance should be adjusted before using binocular microscope 2 Adjustment of diopter As focusing for binocul...

Страница 8: ...are precisely adjusted and approximate clear image should be found with the other objectives after getting the clear image with 10X objective If the image isn t enough clear with the other objectives...

Страница 9: ...water between the objective and specimens will fully grasp the objective function bubbles and impurities can t in immersion oil which will affect image after adjusting with the 40X objective adding mo...

Страница 10: ...ecting the reflector with the base Turning the reflector around the vertical axis so that the reflective side will face to the light direction At the same time around the central axis to rotate the re...

Страница 11: ...cohol and ether proportion 1 4 4 1 2 Alcohol and ether all are burnt easily please take them away from fire Be careful for turn on and off power 4 1 3 Don t clean painted metal and galvanizing metal w...

Страница 12: ...ol Please be sure that the bulb is cool then follow by the operations C Lay aside the microscope reliably unscrew the knurled thumb screw of the lamphold cover plate on the underside of base D Pull ov...

Страница 13: ...screw and pull out the buttom base cover Fig 8 C Replacement Take out LED fastening screw pull out the two power line s connector for LED and replace LED bulb Insert new line s connector separately f...

Страница 14: ...ain the instrument s performance suggest checking instrument regularly If any trouble need to repair please contact with the anufacturer or local dealer 5 Troubleshooting Trouble Causation Remedy Swit...

Страница 15: ...Clear the bubble way Size of iris aperture isn t proper Adjust the size of iris aperture Position of condenser is too low Readjust the position of condenser One side of image is dark or image is movi...