2.

 

Aperture diaphragm adjustment     

Rotating the handle of aperture diaphragm of condenser to adjust diaphragm 

size can adjust the contrast of the specimen.   

In general, aperture diaphragm can be adjusted to 70% or 80% of objective 

exit pupil diameter , will get good image with best contrast.     

3.2.4    Adjustment of interpupillary distance and diopter    (Fig.3) 

1.    Adjustment of Interpupillary Distance 

Please adjusting binocular tube to make left and right field coincide completely . 

▲Interpupillar  distance  is  different  for  everyone,  so  interpupillar  distance 

should be adjusted before using binocular microscope.     

2.    Adjustment of diopter 

As focusing for binocular, user should observe right eyepiece with right eye, and 

make the right eyepiece clear by focusing adjustment, then observe the left 

Fig.3    Adjustment of interpupillary 

distance and diopter 

     

Fig.2

 

Stage and condenser 

 

Clamps swtich knob

 

Specimen 

Stage 
vertical 
adjusting 
knob 

Stage 
horizonal 
adjusting 
knob 

Condenser 
adjusting 
handwheel

 

Aperture 
diaphragm 
lever 

Diopter 
adjusting 
ring 

Interpupillary 
distance dial 

Eyepiece
ec 

Left side 

Right side 

Содержание A11.2601

Страница 1: ...A11 2601 Student Biological Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Страница 2: ...1 CONTENTS Attentions 2 1 Application 3 2 Specifications 4 3 Installation and use 4 4 Maintenance 10 5 Troubleshooting 13...

Страница 3: ...asoline paper and other flammable materials can t near the lamp in case of fire 4 Notes on replacing the bulb Replacement should be based on the identity of the instrument using the same specification...

Страница 4: ...tyle steady structure convenient operation and clear image is suitable for observing various biological specimens and mostly applied in school college hospital etc Binocular head Nosepiece Objective C...

Страница 5: ...Nosepiece Quadruple Condenser Abbe Stage Mechanical stage Fixed stage Optional Focusing Unit Fine adjusting unit division 0 002mm Illumination Mirror Optional 3W LED 12V20W Halogen Optional Accessory...

Страница 6: ...ter with switch used by the instrument is for protecting lamp in case high electric current shock prolong service time 3 2 2 Place sample Fig 2 Please place specimen on the stage and the face with cov...

Страница 7: ...ield coincide completely Interpupillar distance is different for everyone so interpupillar distance should be adjusted before using binocular microscope 2 Adjustment of diopter As focusing for binocul...

Страница 8: ...are precisely adjusted and approximate clear image should be found with the other objectives after getting the clear image with 10X objective If the image isn t enough clear with the other objectives...

Страница 9: ...water between the objective and specimens will fully grasp the objective function bubbles and impurities can t in immersion oil which will affect image after adjusting with the 40X objective adding mo...

Страница 10: ...ecting the reflector with the base Turning the reflector around the vertical axis so that the reflective side will face to the light direction At the same time around the central axis to rotate the re...

Страница 11: ...cohol and ether proportion 1 4 4 1 2 Alcohol and ether all are burnt easily please take them away from fire Be careful for turn on and off power 4 1 3 Don t clean painted metal and galvanizing metal w...

Страница 12: ...ol Please be sure that the bulb is cool then follow by the operations C Lay aside the microscope reliably unscrew the knurled thumb screw of the lamphold cover plate on the underside of base D Pull ov...

Страница 13: ...screw and pull out the buttom base cover Fig 8 C Replacement Take out LED fastening screw pull out the two power line s connector for LED and replace LED bulb Insert new line s connector separately f...

Страница 14: ...ain the instrument s performance suggest checking instrument regularly If any trouble need to repair please contact with the anufacturer or local dealer 5 Troubleshooting Trouble Causation Remedy Swit...

Страница 15: ...Clear the bubble way Size of iris aperture isn t proper Adjust the size of iris aperture Position of condenser is too low Readjust the position of condenser One side of image is dark or image is movi...

Отзывы: