Aperture diaphragm adjustment
Rotating the handle of aperture diaphragm of condenser to adjust diaphragm
size can adjust the contrast of the specimen.
In general, aperture diaphragm can be adjusted to 70% or 80% of objective
exit pupil diameter , will get good image with best contrast.
3.2.4 Adjustment of interpupillary distance and diopter (Fig.3)
1. Adjustment of Interpupillary Distance
Please adjusting binocular tube to make left and right field coincide completely .
▲Interpupillar distance is different for everyone, so interpupillar distance
should be adjusted before using binocular microscope.
2. Adjustment of diopter
As focusing for binocular, user should observe right eyepiece with right eye, and
make the right eyepiece clear by focusing adjustment, then observe the left