background image

eyepiece, at the same time, adjust the diopter ring of the left eyepiece 

tube(diopter difference compensation)to make the image of left eyepiece clear as 

same as the right eyepiece. 

3.2.5    Focusing unit

Fig.4

 

Usually 10X objective with large field of view and long depth of field should 

be used for focusing adjustment first.   

Take  10X  objective  into  bright  path,  then  observe  the  specimen  with  right 

eyepiece by right eye and turn coarse focusing knob to find the image, then turn 

slowly fine focusing knob to make the image clear. 

All  objectives  of  our  microscope  are  precisely  adjusted,  and  approximate 

clear  image  should  be  found  with  the  other  objectives  after  getting  the  clear 

image  with  10X  objective.  If  the  image  isn’t  enough  clear  with  the  other 

objectives, please turn fine focusing knob slightly to find the satisfactory image. 

 

 

 

 

 

 

   

▲Don’t  rotate  the  left  &  right  coarse  and  fine  focusing  knobs  in  reverse 

direction at the same time, if so, the focusing system will be damaged. 

▲Don’t directly pull objective to turn nosepiece when user changes the different 

objectives,if so, optical quality of microscope possibly be affected.The right way 

Fig. 4 Coarse & fine focusing knobs and tension adjusting knobs 

 

 

Right coarse 
focusing knob

 

Left fine 
focusing knob 

Left coarse 
focusing 
knob   

Tension 
adjusting knob 

Right fine 
focusing knob 

Содержание A11.2601

Страница 1: ...A11 2601 Student Biological Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Страница 2: ...1 CONTENTS Attentions 2 1 Application 3 2 Specifications 4 3 Installation and use 4 4 Maintenance 10 5 Troubleshooting 13...

Страница 3: ...asoline paper and other flammable materials can t near the lamp in case of fire 4 Notes on replacing the bulb Replacement should be based on the identity of the instrument using the same specification...

Страница 4: ...tyle steady structure convenient operation and clear image is suitable for observing various biological specimens and mostly applied in school college hospital etc Binocular head Nosepiece Objective C...

Страница 5: ...Nosepiece Quadruple Condenser Abbe Stage Mechanical stage Fixed stage Optional Focusing Unit Fine adjusting unit division 0 002mm Illumination Mirror Optional 3W LED 12V20W Halogen Optional Accessory...

Страница 6: ...ter with switch used by the instrument is for protecting lamp in case high electric current shock prolong service time 3 2 2 Place sample Fig 2 Please place specimen on the stage and the face with cov...

Страница 7: ...ield coincide completely Interpupillar distance is different for everyone so interpupillar distance should be adjusted before using binocular microscope 2 Adjustment of diopter As focusing for binocul...

Страница 8: ...are precisely adjusted and approximate clear image should be found with the other objectives after getting the clear image with 10X objective If the image isn t enough clear with the other objectives...

Страница 9: ...water between the objective and specimens will fully grasp the objective function bubbles and impurities can t in immersion oil which will affect image after adjusting with the 40X objective adding mo...

Страница 10: ...ecting the reflector with the base Turning the reflector around the vertical axis so that the reflective side will face to the light direction At the same time around the central axis to rotate the re...

Страница 11: ...cohol and ether proportion 1 4 4 1 2 Alcohol and ether all are burnt easily please take them away from fire Be careful for turn on and off power 4 1 3 Don t clean painted metal and galvanizing metal w...

Страница 12: ...ol Please be sure that the bulb is cool then follow by the operations C Lay aside the microscope reliably unscrew the knurled thumb screw of the lamphold cover plate on the underside of base D Pull ov...

Страница 13: ...screw and pull out the buttom base cover Fig 8 C Replacement Take out LED fastening screw pull out the two power line s connector for LED and replace LED bulb Insert new line s connector separately f...

Страница 14: ...ain the instrument s performance suggest checking instrument regularly If any trouble need to repair please contact with the anufacturer or local dealer 5 Troubleshooting Trouble Causation Remedy Swit...

Страница 15: ...Clear the bubble way Size of iris aperture isn t proper Adjust the size of iris aperture Position of condenser is too low Readjust the position of condenser One side of image is dark or image is movi...

Отзывы: