NOTICE
1. Read all of the information in the operation guide and any other literature in the box before
operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same wrist for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood
circulation in your arm to recover. Prolonged over-inflation (cuff pressure exceed 300
mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause
ecchymoma of your arm.
7. Consult your physician if you have any doubt about below cases:
1) The application of the cuff over a wound or inflammation diseases;
2) The application of the cuff on any limb where intravascular access or therapy, or an arterio-
venous (A-V) shunt, is present;
3) The application of the cuff on the arm on the side of a mastectomy or lymph node
clearance;
4) Simultaneously used with other monitoring medical equipments on the same limb;
5) Need to check the blood circulation of the user.
8. This Electronic Sphygmomanometers is designed for adults and should never be used on
infants or young children. Consult your physician or other health care professionals before use
on older children.
9. Do not use this unit in a moving vehicle, This may result in erroneous measurement.
10. Blood pressure measurements determined by this monitor are equivalent to those obtained
by a trained observer using the cuff/stethoscope auscultation method, within the limits
prescribed by the American National Standard Institute, Electronic or automated sphygmo-
manometers.
11. Information regarding potential electromagnetic or other interference between the blood
pressure monitor and other devices together with advice regarding avoidance of such inter-
ference please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION.It is suggested that
the blood pressure monitor be kept 30 cm away from other wireless devices, such as WLAN
unit, cell phone, microwave oven, etc. It can’t be used near active HF SURGICAL EQUIPMENT
and the RF shielded room of an ME SYSTEM for magneticresonance imaging, where the in-
tensity of EM DISTURBANCES is high.
12. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of
blood pressure measurement, a signal of
(
( )
)
will be displayed. Under this condition, the
Electronic Sphygmomanometer can keep function, but the results may not be accurate, it’s
suggested that you consult with your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed: