background image

Содержание OP-904

Страница 1: ...optimascale com OP 904 PANEL MOUNT INDICATOR USER S MANUAL ...

Страница 2: ...ions 3 Display and Key Descriptions 4 Operating Instructions 5 Rear Panel 6 General Functions 7 Function Settings 8 Parameter Settings 9 10 Calibration Settings 11 14 Check Weighing Configuration 15 Cable Connection Manual 16 18 Troubleshooting 19 Contact Us 19 ...

Страница 3: ...any of the components are cracked DO NOT use for purposes other then weight taking To avoid damaging the battery do not keep charger plugged in once battery is fully charged Make sure the weight is not over the Max capacity as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Anot...

Страница 4: ...ix wire Keyboard six key Display six digits red 7 segment LED display Relay output 4 output AC250v 5A Input isolation voltage 2500V Analog output 4 20mA 0 5v Serial port RS232 RS485 baud rate 600 19200bit s Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70 C 32 104 F Housing dimension 92 x 45mm Front panel dimension 102 x 55mm Trepanning dimension 93 x 46mm Softw...

Страница 5: ...3 SPECIFICATIONS FIGURE 1 INDICATOR MEASUREMENTS 2 16 4 5 46 1 78 5 6 6 1 ...

Страница 6: ...t you have zero d the scale Motion The weight on the scale is unstable Gross Shows you are in Gross weight mode includes tare default mode Net Shows you are in Net weight mode weight without tared weight kg The weight is shown in kilograms t The weight is shown in tons lb The weight is shown in pounds Power Flashes red low battery Solid red charging Green fully charged Over Flashes when weight is ...

Страница 7: ...u will receive an error and will need to tare out the weight Tare Function The Tare function is used when you only wish to see the current change in weight not the entire amount of weight that is on the scale When the indicator is in gross mode gross light is shown pressing the TARE key will Tare the current weight on the scale and enter the net mode net light shown For example if you are using a ...

Страница 8: ...D GND for RS232 A B for RS485 Analog output Vout Iout GND Relay output J1 J1 J2 J2 J3 J3 J4 J4 Input YX1 YX2 YX3 YX4 YXCOM 0vTo exit Tare mode press the TARE key again to enter gross mode and you will see the total weight of the container and the product Note If you remove the container the scale will show the minus weight of the container ...

Страница 9: ...ight mode press 1 FinAL 5 Enter to test mode Turn power on press and 1 dsp 7 1 Reset all parameters back to default Turn power on turn the calibration switch to ON press and i ALL 7 2 Reset general function parameters back to default Turn power on press and 1 FnC 7 2 Parameter Settings Key meaning Key Meaning Save and Exit Move the flashing digit to the left Move the flashing digit to the right In...

Страница 10: ...for serial port interface Press to display 03 Sq for weight comparison procedures Press to display 04 AnL for analog current output Press to display 05 In for external input interface Press to enter the next menu level Press to save and exit out of this menu and restart 01 FnC 01 FnC 01 FnC 01 FnC 01 FnC FnC 01 In 01 AnL 01 Sq 01 232 01 Weighing Mode and ...

Страница 11: ... Default Digital Filter 1 FnC 01 00 01 10 Greater Less 03 Digital Filter 2 FnC 02 00 01 02 03 Greater Less 02 Rate for display rewrite FnC 03 01 05 10 20 120 1 times sec 5 times sec 10 times sec 20 times sec 120 times sec 10 02 232 Serial Port Interface Function Item Parameter Description Default Baud Rate 232 01 00 01 02 03 04 05 600bit s 1200bit s 2400bit s 4800bit s 9600bit s 19200bit s 04 Tran...

Страница 12: ...Function Item Parameter Description Default Signal output AnL 01 00 01 02 4 20mA current 0 20mA current 0 5v voltage 00 Data format AnL 02 00 01 display weight gross weight 00 Maximum weight AnL 03 010000 when max analog output weight 010000 Zero adjust AnL 04 4mA 0mA 0v Press or to adjust output 4 Linearity adjust AnL 05 20mA 5v Press or to adjust output 20 Output range limit AnL 06 00 01 no limi...

Страница 13: ...ibration 3 Press to display 02 CAL General calibration Follow the steps to calibrate your scale 4 Press to display 03 CUo Sensitivity calibration Follow the steps to set filters and manage the sensitivity of your scale 5 When you are done with calibrating make sure to turn the switch to OFF 6 The screen will display End save the data and restart 02 CAL Calibration 01 CSP Calibration Set up 03 CUo ...

Страница 14: ...5 division size 02 Max Capacity CSP 04 010000 Max capacity 1000 Zero Setting range CSP 05 00 01 02 0 1 2 02 Initial zero setting range CSP 06 00 01 02 05 10 0 1 2 5 10 10 Automatic zero setting range CSP 07 00 05 10 20 0 0 5d 1d 2d 05 Automatic zero setting time CSP 08 00 01 02 03 0 1 sec 2 sec 3 sec 01 Stable time CSP 09 00 01 02 fast medium slow 01 Stable range CSP 10 01 02 05 10 1d 2d 5d 10d 02...

Страница 15: ...o calibration press Display zero code 3 seconds Weight calibration press Skip weight calibration press Use to set the weight Once the weight is stable press Sensitivity 3 seconds calibration done Set calibration switch to OFF You are finished Note Zero calibration can be no more than 5mv 01 CSP 02 CAL ZERO 288888 SPAN 01000kg 120000 02 CAL END ...

Страница 16: ...ather zero code Or press to set zero code to enter zero calibration Sensitivity calibration press Skip sensitivity calibration press Use to set the sensitivity enter press Sensitivity calibration done Set calibration switch to OFF You are finished Sensitivity 100000 x load cell full scale output x weigh ing meter capacity N x Load cell capacity 01 CSP 03 CUo ZERO 288888 SPAN 120000 02 CUo END ...

Страница 17: ...F FALL Free fall value 6 oVer Over value 7 UndEr Under value 8 Z bAnd Zero band Warning set FinAL SP1 SP2 SP3 F Fall set Sq01 1 net weigh Signal Output Condition Relay Output SP1 Net Final SP1 J1 SP2 Net Final SP2 J2 SP3 Net Final SP3 J3 Free Fall Net Final F FALL J4 Set Sq01 3 Signal Output Condition Relay Output HI Net SP1 J1 OK SP1 Net SP2 J2 LO Net SP2 J3 ...

Страница 18: ...C SEN EXC SEN SIG SIG NOTE use 4 wire load cell need EXC and SEN short connect EXC and SEN short connect Serial Port TXD RS232 transmission RXD RS232 receive GND RS232 GND A RS485 A B RS485 B Analog Output Vout 0 5v voltage load more than 1kΩ Iout 4 20mA current load span 100 500Ω GND GND ...

Страница 19: ...up relay output J2 J2 second group relay output J3 J3 third group relay output J4 J4 fourth group relay output Relay Output Connecction diagram Common connection diagram Note Use common connection put one of J1 J2 J3 J4 short connect ...

Страница 20: ... group input YX3 third group input YX4 fourth group input YXCOM input common 0v out connect power GND Input Diagram Out connect power input diagram Note Using out elctric power needs more than 3V battery charge Out electric power less than 24V ...

Страница 21: ...e correct weight within the defined range ERR2 During calibration the weight is below the minimum required weight The calibration weight minimum is 10 of the max capacity set in C04 Recommended to use 60 80 of max capacity if possible ERR3 During calibration the input signal is negative 1 Check all wire connections 2 Check load cell 3 Recalibrate 4 PCB replacement needed if steps 1 3 fail ERR4 Dur...

Страница 22: ......

Отзывы: