9
INDICATOR PARAMETER SETTINGS
To enter parameter settings, follow the procedure below:
1. Press ENTER
and MODE
at the same time for 2 seconds to enter the function settings
2. Navigate through the settings as shown in the table below by using the arrow keys and return
keys as labeled under each indicator button
3. Press the ENTER
key to enter/edit the parameter setting
Press the
key to save and exit settings at any time
01 FnC Function Parameter Code
Function
Item
Parameter
Description
Default
Digital Filter 1
FnC 01
00
01
...
10
Greater
Less
03
Digital Filter 2
FnC 02
00
01
02
03
Greater
Less
02
Rate for display
rewrite
FnC 03
01
05
10
20
120
1 times/sec
5 times/sec
10 times/sec
20 times/sec
120 times/sec
10
02 232 Serial Port Interface
Function
Item
Parameter
Description
Default
Baud Rate
232 01
00
01
02
03
04
05
600bit/s
1200bit/s
2400bit/s
4800bit/s
9600bit/s
19200bit/s
04
Transmit Mode
232 02
00
01
02
03
04
close serial port
continuous
print mode
stable send
command mode
01
Data Format
232 03
00
01
02
format 1
format 2
print format
00
Transmit Time
232 04
00
01
02
03
open
1 sec
2 sec
5 sec
00
Address for Multi-
computer
232 05
00
01~99
Only one
Address (bcd)
00