8.0 TROUBLESHOOTING

6. One side of the image 

is out of focus.

The nosepiece is not in the cen-

ter of the light path

Turn the nosepiece to a click stop

The specimen is out of place 

(tilted)

Place the specimen flat on the stage.

The optical performance of the 

sample cover glass is poor

Use a cover glass of better quality

II. Mechanical section

1.The coarse focus knob 

is hard to turn.

The tension adjustment collar is 

too tight

Loosen the tension adjustment collar

2.The focus is unstable.

The tension adjustment collar is 

too loose

Tighten the tension adjustment collar

III. Electric section

1. The LED doesn’t turn 

on.

No power supply

Check the power cord connection

2. The brightness is not 

enought

The brightness adjustment is low Adjust the brightness

3. The light blinks

The power cord is poorly con

-

nected

Check the power cord

IV. Viewing tube assembly

The field of view of the 

two eyes is different.

The interpupillar distance is not 

correct

Adjust the interpupillar distance

The dioptric correction is not 

right

Adjust the dioptric correction

The viewing technique is not cor

-

rect, and the operator is straining 

the eyesight

When look into the objective, do not stare at 

the specimen but look at the whole field of 

view. Periodically, move the eyes away to 

look at a distant object, then back into the 

objective

V. Microphotography and video

1. The image is unfocu

-

sed.

Incorrect focusing

Adjusting the focus system as in the present 

manual

2. The edge of the image 

is unfocused.

To some degree, it is inherent to 

the nature of achromatic objec

-

tives

The problem can be minimized by a correct 

setting of the aperture diaphragm

3. Bright patches appear 

on the image.

Stray light is entering the micro

-

scope through the eyepieces 

and through the camera viewfin

-

der

Cover the eyepieces and the viewfinder with 

a dark cloth

Содержание XDS-3

Страница 1: ...OPTIKA MICROSCOPES ITALY www optikamicroscopes com info optikamicroscopes com Ver 4 0 0 XDS 3 OPERATION MANUAL GUIDA UTENTE...

Страница 2: ...S page 6 3 0 INSTALLATION page 7 4 0 ADJUSTING THE MICROSCOPE page 11 5 0 PHASE CONTRAST OPERATION page 15 6 0 MICROPHOTOGRAPHY AND VIDEO page 16 7 0 TECHNICAL SPECIFICATIONS page 17 8 0 TROUBLESHOOTI...

Страница 3: ...o industry standards for safety instrumentation laboratory Misuse can cause personal injury or damage to the instrument Keep this manual at hand close to the instrument for an easy consultation Electr...

Страница 4: ...1 main switch MAINTENANCE 1 Use a lint free cloth to wipe the glass parts In order to remove fingerprints and oil stains slightly dampen the cloth with xylene or with a 3 7 mixture of ethanol and ethe...

Страница 5: ...AFETY SYMBOLS Symbol Meaning The surface may be hot do not touch Please read the instruction carefully before use Improper operation will result in injuries or malfunction Main switch ON position Main...

Страница 6: ...ves 10X PH 20X PH 40X Phase contrast slider LED housing Illumination bracket Condenser Stage insert Objective turret Nosepiece fixed Stage fixed Stage extension Glass stage Microscope base Trinocular...

Страница 7: ...Before installing be sure every component is clean Take care not to scratch parts or surfaces Store the supplied hexagon wrench in a safe place When changing the components you will need it again Mic...

Страница 8: ...port the nosepiece is placed in the lo west position and the tension adjustment collar 2 is adjusted to the appropriate tension when the micro scope leaves the factory 2 Screw the lowest magnification...

Страница 9: ...the glass stage 1 make sure that the stage is horizontal 2 Install the stage insert in the stage opening Turn the disk until the V groove faces the user This simplifies objective identification 3 2 5...

Страница 10: ...heck for a safe connection Please use the supplied power cord If lost or damaged please refer to qualified service Connect the power cord to a grounded earthed power supply only 3 2 8 Replacing the fu...

Страница 11: ...HE MICROSCOPE Dioptric adjustment ring Light path selector lever Tension adjustment collar Coarse focus knob Fine focus knob Phase contrast slider Phase centering bolt Aperture diaphragm adjustment le...

Страница 12: ...the tension 4 2 STAGE 4 2 1 Setting the specimen fig 15 16 For the best image quality use flasks Petri dishes and slides with a 1 2 mm thickness Using 35mm culture dishes You can place a 35 mm cultur...

Страница 13: ...espond to the ope rator s dioptric correction 4 3 2 Adjusting the interpupillary distance fig 18 Observing with both eyes hold the two eyepiece prism assemblies Rotate them around their common axis un...

Страница 14: ...identify the aperture diaphragm the eyepiece can be re moved or the centering telescope can be used The aperture diaphragm looks as in fig 21 The aperture can be changed by moving the aperture adjust...

Страница 15: ...ig 25 Usually this operation is not needed If necessary please pro ceed with the following steps 1 Place a specimen on the stage and focus it 2 Take out the eyepiece from the tube without the dioptric...

Страница 16: ...alling the photography adapter fig 28 1 Loosen the locking bolt 1 on the trinocular viewing tube and take out the dust cap 2 2 Install the photography adapter into the tri through port accor ding to i...

Страница 17: ...ll plates etc not included in LT version Pair of side extensions to expand the surface of the stage Glass stage insert with hole for small dimension specimens Nosepiece Quintuple revolving nosepiece r...

Страница 18: ...dust on the eyepiece Clean the eyepiece 4 There is an apparent double image The size of the aperture diaphragm is too small Open the aperture diaphragm 5 Poor image quality The image is not sharp The...

Страница 19: ...ted Check the power cord IV Viewing tube assembly The field of view of the two eyes is different The interpupillar distance is not correct Adjust the interpupillar distance The dioptric correction is...

Страница 20: ...waste The separate collection of this equipment at the end of its lifetime is organized and managed by the producer The user will have to contact the manufacturer and follow the rules that he adopted...

Страница 21: ......

Страница 22: ...26 3 0 INSTALLAZIONE pagina 27 4 0 REGOLAZIONE DEL MICROSCOPIO pagina 31 5 0 OSSERVAZIONI IN CONTRASTO DI FASE pagina 35 6 0 MICROFOTOGRAFIA E VIDEO pagina 36 7 0 SPECIFICHE TECNICHE pagina 37 8 0 RIS...

Страница 23: ...aboratorio L utilizzo non corretto pu causare lesioni alla persona o danni allo strumento Mantenere questo manuale a portata di mano vicino allo strumento per una facile consultazione Precauzioni di s...

Страница 24: ...te il cavo di rete fornito in dota zione 2 3 1 Interruttore principale MANUTENZIONE 1 Per pulire le parti in vetro si utilizzi un panno morbido Per rimuovere impronte digitali o macchie di unto inumid...

Страница 25: ...mbolo Significato La superficie potrebbe scottare Non toccare Leggere attentamente le istruzioni prima dell uso Un utilizzo scorretto potrebbe portare a lesioni o malfunzionamento Interruttore princip...

Страница 26: ...ider per contrasto di fase Allogiamento LED Braccio porta illuminatore Condensatore Inserto piano Torretta portaobiettivi Revolver fissa Piano fisso Prolungamento del piano Piano in vetro Base del mic...

Страница 27: ...i prima di procedere all installazione Assicurarsi che non vi siano graffi sulle componenti ottiche n sulle superfici Riporre la chiave esagonale fornita in dotazione in un posto sicuro perch sar nece...

Страница 28: ...urante il trasporto prima della spedizione la torretta viene messa nella posizio ne pi bassa e si sistema l anello di regolazione della tensione 2 nella tensione appropriata 2 Avvitare l obiettivo con...

Страница 29: ...tapreparati sia perfettamente orizzontale quando si usa il piano in vetro 1 2 Inserire l inserto in vetro nell apertura del piano Girare il disco finch la scanalatura a V si trova di fronte all operat...

Страница 30: ...a sicurezza del collegamento Si utilizzi il cavo fornito in dotazione Se viene perso o dan neggiato si contatti il servizio assistenza Il cavo va collegato soltanto a una presa di corrente con messa a...

Страница 31: ...luminoso Anello di regolazione della tensione Manopola di regolazione della messa a fuoco macrometrica Manopola di regolazione della messa a fuoco micrometrica Slider per contrasto di fase Bullone di...

Страница 32: ...ontrario 4 2 PIANO PORTAPREPARATI 4 2 1 Inserimento del preparato fig 15 16 Per ottenere la migliore qualit delle immagini si consi glia l uso di beute capsule Petri e vetrini con uno spes sore di 1 2...

Страница 33: ...pondere alla correzione diottrica dell operatore 4 3 2 Regolazione della distanza interpupillare fig 18 Osservando con entrambi gli occhi sostenere il gruppo di oculari Ruotare questi lungo l asse com...

Страница 34: ...oppure utilizzare il telescopio di centratura Il diafram ma di apertura appare come in fig 21 T L apertura pu essere variata operando sull apposita levetta di regolazione 1 l immagine del diaframma d...

Страница 35: ...aperto 5 2 2 Centratura del cerchio luminoso fig 25 Solitamente non necessario effettuare questa operazione Nel caso lo fosse seguire la procedura descritta di seguito 1 Posizionare un preparato sul...

Страница 36: ...l adattatore foto fig 28 1 Svitare il bullone di bloccaggio 1 sul tubo trinoculare e ri muovere il tappo antipolvere 2 2 Installare l adattatore foto nella porta trinoculare seguendo le istruzioni qui...

Страница 37: ...versione LT Coppia di estensioni laterali per aumentare la superficie di appoggio del tavolino Piattello forato in vetro per alloggiare campioni di piccole dimensioni Revolver Revolver portaobiettivi...

Страница 38: ...lare ci sono polvere e macchie Pulire l oculare 4 L immagine appare doppia Il diaframma di apertura troppo chiuso Aprire il diaframma di apertura 5 La qualit delle imma gini scarsa L immagine non niti...

Страница 39: ...collegato bene Verificare il collegamento del cavo IV Montaggio oculari Il campo visivo diverso per ciascun occhio La distanza interpupillare non corretta Regolare la distanza interpupillare La correz...

Страница 40: ...tura giunta a fine vita organizzata e gestita dal produttore L utente che vorr disfarsi della presente apparecchiatura dovr quindi contattare il produttore e seguire il sistema che questo ha adottato...

Страница 41: ......

Страница 42: ...ientificos c Puig i Pidemunt n 28 1 2 Pol Ind Pl d en Boet 08302 MATARO Barcelona Espa a Tel 34 937 586 245 34 937 414 529 New York Microscope Company Inc 100 Lauman Lane Suite A Hicksville New York 1...

Отзывы: