Page 6
9. Maintenance
Microscopy environment
This microscope is recommended to be used in a clean, dry and shock free environment with a temperature of
5°-40°C and a maximum relative humidity of 75 % (non condensing). Use a dehumidifier if needed.
To think about when and after using the microscope
• The microscope should always be kept vertically when moving it and be careful so that no
moving parts, such as the eyepieces, fall out.
• Never mishandle or impose unnecessary force on the microscope.
• Never attempt to service the microscope yourself.
• After use, turn off the light immediately, cover the microscope with the included
dust-cover, and keep it in a dry and clean place.
Electrical safety precautions
• Before plugging in the power supply, make sure that the supplying voltage of your region
matches with the operation voltage of the equipment and that the lamp switch is in off-
position.
• Users should observe all safety regulations of the region. The equipment has acquired
the CE safety label. However, users do have full responsibility to use this equipment safely.
Cleaning the optics
• If the optical parts need to be cleaned try first to: use compressed air.
• If that is not sufficient: use a soft lint-free piece of cloth with water and a mild detergent.
• And as a final option: use the piece of cloth moistened with a 3:7 mixture of ethanol and ether.
Note: ethanol and ether are highly flammable liquids. Do not use them near a heat source, near sparks or
near electric equipment. Use these chemicals in a well ventilated room.
• Remember to never wipe the surface of any optical items with your hands. Fingerprints can damage the
optics.
• Do not disassemble objectives or eyepieces in attempt to clean them.
For the best results, use the OPTIKA cleaning kit (see catalogue).
If you need to send the microscope to Optika for maintenance, please use the original packaging.
10. Troubleshooting
Review the information in the table below to troubleshoot operating problems.
PROBLEM
CAUSE
SOLUTION
1. Optical System
LED does not light.
Power cord is unplugged.
Plug power cord into the power outlet.
The main switch is on “O” position
Set main switch on “I” position.
Light too dim on the sample
Intensity set too low.
Adjust the intensity knob.
Содержание CL-14
Страница 1: ...Model CL 14 Accessories Series INSTRUCTION MANUAL v 1 0 2017 ...
Страница 8: ......
Страница 9: ...Modello CL 14 Serie Accessori MANUALE D ISTRUZIONI v 1 0 2017 ...
Страница 16: ......
Страница 17: ...Modelo CL 14 Serie accesorios MANUAL DEL USUARIO v 1 0 2017 ...
Страница 24: ......
Страница 25: ...Modèle CL 14 Série accessoires MANUEL D INSTRUCTIONS v 1 0 2017 ...
Страница 32: ......
Страница 33: ...Modell CL 14 Zubehörteile Serien BEDIENUNGSANLEITUNG v 1 0 2017 ...