39
Relay Setting
Item
Description
Refer to
Relay Setting The following settings are available.
Alarm output setting:
Configure various settings to output an alarm when
the measurement value exceeds a fixed value. The
measurement item set on "Display Item" P.22 can be
selected as the output item of alarm output.
Maintenance output setting:
Configure the settings to notify the user when maintenance
will be performed.
<Setup Workflow>
1. Configure the setting of "Maintenance Timer" P.70.
2. Configure the setting of "Maintenance output setting"
Sensor cap output setting:
Set to notify the replacement time of the sensor cap.
<Setup Workflow>
Refer to the sensor's instruction manual to configure the
setting in the following order.
1. Configure the setting of "Sensor Cap Replacement
Timer".
2. Configure the setting of "Sensor Cap Output Setting".
Sensor's
instruction
manual
Cleaning output setting:
The operation of the cleaning device can be set using relay
output.
<Setup Workflow>
1. Configure the setting of "Cleaning Interval (Relay)" P.76.
2. Configure the setting of "Cleaning Activation" P.77 as
needed.
3. Configure the setting of "Post-cleaning Standby" P.78
as needed.
4. Configure the setting of "Cleaning output setting" P.47.
Self Checking output setting:
Set to notify malfunctions such as sensor failure and
disconnection.
When Self Checking output is selected, NO and NC are
reversed.
Relay Test
Mode
Switch the relay ON/OFF to check the output status to the
control panel, etc.