FluxGage User Manual
19
7.2
Control Parameters
7.2.1
Autoset
‘Autoset’ sets the optimal spectrometer integration time and other internal settings by
measuring the incoming light from a DUT.
7.2.2
Background
‘Background’ should be done after Autoset with the DUT in place and turned off. The FluxGage
stores the background value of the spectrometer and later subtracts it from the actual
measurement.
7.2.3
Averages
Adding averaging will improve the stability of the measurement. If the integration time
multiplied by the number of averages exceeds 500ms, the photovoltaics and photodiode will
also increase their measurement time and improve their stability.
7.2.4
Abort
The abort button resets the current measurement and starts a new one. This is useful if the
current measurement take a long time, for example due to many averages.
7.2.5
More settings
The ‘more setting’ button extends the SW window to reveal more options described in the next
section.