FluxGage User Manual
38
In the setting area, enter 0 in the luminous flux box
Click ‘Open Light File’ and select the spectral calibration file of the calibration source.
If the file is loaded correctly, the ‘calibration file valid’ indicator will turn green.
Turn the calibration source on.
Press ‘Autoset’ and wait for the operation to complete.
In the setting area, select 30 averages.
Turn off the calibration source.
Press ‘Background’ and wait for the operation to complete.
Turn the calibration source on and allow enough time to reach thermal stabilization
(~30minutes).
Press ‘Calibrate’ and wait for the operation to complete.
At the top of the screen, the measured flux of the calibration source will appear. It
should match the value entered in the settings area.
Click on the top right corner to exit the window.
8.4
Back up calibration
In addition to the working calibration there is a backup calibration.
The backup calibration can be retrieved by pressing ‘Restore Backup Value’. This will copy the
backup calibration into the working calibration. The backup calibration can be overwritten by
pressing ‘Overwrite Backup Value’. This should be done when the calibration is done with an
Ophir certified calibration unit.
Rev 1.02-3/01.01.18/ds – updated sections 7.8,
7.9, 7.14, 7.15 (see details in email)
Rev 1.02-2/01.01.18/ds - updated
Rev 1.02-1/21.12.16/er added abort
Rev 1.00-1/28.09.16/er - new
FluxGage User Manual
24 July, 2018
Rev 2.0.15-1
Содержание FluxGage
Страница 1: ...FluxGage LED Luminaire Measurement System User Manual FluxGage FluxGage...
Страница 12: ...FluxGage User Manual 12 Click on Next...
Страница 13: ...FluxGage User Manual 13 Click on Next...
Страница 14: ...FluxGage User Manual 14 Install in progress...
Страница 17: ...FluxGage User Manual 17 7 Operating the FluxGage SW...
Страница 33: ...FluxGage User Manual 33...
Страница 34: ...FluxGage User Manual 34...
Страница 35: ...FluxGage User Manual 35...