![Ophir-Spiricon IBP-YAG Series Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/ophir-spiricon/ibp-yag-series/ibp-yag-series_user-manual_1640325021.webp)
Cooling Water Requirements
Appendix A
Cooling Water Requirements
The IBP-YAG needs a reliable flow of cooling water to dissipate the energy of the laser. The
required cooling flow is dependant on the laser power level and the initial coolant temperature.
Note: If the inlet cooling water is chilled to below the dew point, condensation will occur inside
the unit. To prevent water damage to the measurement head, purge air must be provided
DO NOT USE DE-IONIZED (DI) WATER. DI WATER
WILL QUICKLY CAUSE SEVERE CORROSION OF ALL
COMPONENTS THAT CONTACT THE WATER.
Water Pressure not to exceed 200 kPa (30 psi).
To maintain a safe working environment, do not allow the outlet water temperature to exceed
49°C (120
o
F).
A-1
The chart and table below indicate cooling flow requirements for various laser power levels.
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
0
1
2
3
4
5
6
Water Flow (GPM)
O
u
tl
et Temperature Ri
se (°F)
6 kW
4 kW
2 kW
1 kW
IBP-YAG