Ophir-Spiricon IBP-YAG Series Скачать руководство пользователя страница 16

Maintenance 

and 

Storage

Chapter 3 

 

Maintenance and Storage 

 

3-1 

 

3.1

 

Storing and transporting the IBP-YAG 

If the IBP-YAG will be stored for an extended time, transported, or will be exposed to freezing 

temperatures, the cooling water must be drained to avoid damage. 
Disconnect the electrical and signal cables from the IBP-YAG.  Disconnect the Cooling water 

lines and drain the water into a collection basin.  To completely drain the water from the 

instrument, it will be necessary to invert the instrument so that the water outlet is lower than 
the inlet.   

3.2

 

Cleaning and Replacing Optics 

The Primary sampling mirror must be kept clean and dust free to avoid damaging the coating.  Any 
contamination will increase the absorption of laser energy and lower the damage threshold when 

exposed to a high power beam 

To reduce the risk of damage to the IBP-YAG optics 

and minimize the amount of cleaning required, keep 

the shutter closed whenever not actually taking 

measurements.  Using Purge Air in the IBP-YAG can 

also reduce the amount of dust that collects on the 

sampling optics. 

3.2.1

 

Cleaning the sampling optics 

To remove small amounts of dust from the primary sampling mirror, open the shutter and use 
photographic type canned air and lens brushes to carefully clean the sampling mirror.  Use care to 

avoid transferring dust deeper into the instrument.  Using purge air during this procedure will 

greatly reduce the possibility of transferring dust to other optical elements.   

Avoid using shop air to clean optics because it often 

contains water and/or oil.  These can create 

absorbing films on the optics and destroy the 

coatings. 

If this procedure does not remove the debris, the unit will need to be partially disassembled to 

remove and clean the optics.  Follow the procedure in the next section to disassemble and remove 
the primary optics for cleaning. 
Great Care should be taken when handling optics.  Observe the following precautions to reduce the 

chance of damage: 

 

Wear powder free rubber/latex gloves or finger cots when handling optics.  Skin oils can 
contaminate the coatings and may cause permanent damage. 

 

Always place optics on lens tissues for protection. 

 

never place optics on hard or rough surfaces.  The coatings are easily scratched. 

 

Use high quality lens tissues.  The use of reagent grade acetone and isopropanol is 
recommended. 

IBP-YAG 

Содержание IBP-YAG Series

Страница 1: ...W 1000 N Main Logan UT 84321 For Sales Service or Technical Support Phone 435 753 3729 Fax 435 753 5231 Email service ophir spiricon com Email sales ophir spiricon com Copyright 2008 Ophir Spiricon In...

Страница 2: ...manual at any time and without notice All rights to the product and any accompanying operators manuals are reserved While every precaution has been taken in the preparation of this product the publis...

Страница 3: ...__________ 2 1 2 3 Frame Grabber Setup ________________________________________2 2 2 3 1 Refer to your LBA PC Operators Manual ___________________________ 2 2 2 3 2 Starting LBA PC for the first time_...

Страница 4: ...and use procedures We also suggest that you read the software manual to become familiar with the controls and the measurements provided Chapter 1 contains general Information and specifications of the...

Страница 5: ...ndensing Note When storage temperature may drop below 5 C all cooling water must be removed from the unit to prevent damage due to freezing Operating Temperature 10o C to 50o C 50 F to 122 F Operating...

Страница 6: ...ons Weight 12 kg 26 lb Dimensions 24 x 50 x 32 cm 9 6 x 19 5 x 12 6 17 8cm 7 0 32 1cm 12 6 22 8cm 9 0 26 0cm 10 2 48 4cm 19 1 24 3cm 9 6 19 7cm 7 8 45 7cm 18 0 Figure 1 1 Overall Dimensions IBP YAG Ge...

Страница 7: ...to connect the IBP YAG to the LBA 7xx frame grabber card 3 Cooling Water Inlet 4 Cooling Water Outlet These fittings are standard barbed hose connectors for 5 16 ID Hose Refer to appendix A for water...

Страница 8: ...e shutter closed except when the laser is directed into this opening 9 Shutter Release Lift up on the top handle to close the shutter 10 Bubble Level 11 Leveling Adjusters Use the leveling adjusters t...

Страница 9: ...tive eyewear when operating the laser Do not set up the IBP YAG or any optical components at eye level Do not wear reflective jewelry or watches while operating this equipment to reduce the possibilit...

Страница 10: ...nal Cart system If ordered the cart system will be shipped separately and will include separate unpacking and setup instructions 2 2 Setting up the IBP YAG 2 2 1 Frame Grabber and Software Setup The I...

Страница 11: ...this connection remove the plug from the back of the instrument and replace it with an 1 8 NPT hose fitting Connect to a supply of clean dry nitrogen or air at about 5 psi 2 2 2 3 Power and Video conn...

Страница 12: ...dialog box and under Method select the Continuous option Then click OK If there still is no indication of operation verify in the lower status bar that the Video V enunciator is illuminated Green If t...

Страница 13: ...tions or when using chilled cooling water run the purge air for 10 min before starting the cooling water Set fine attenuation knob to the minimum position by turning the knob clockwise until it stops...

Страница 14: ...saturated portions of the displayed beam will turn white when saturated Increase the fine attenuation until beam is just below saturation Center the Beam image in the display by adjusting the levelin...

Страница 15: ...r s shutter or by turning the laser off Never attempt to block the laser beam by closing the IBP YAG shutter The IBP YAG shutter is not designed to handle high energy beams and doing so will damage th...

Страница 16: ...cs 3 2 1 Cleaning the sampling optics To remove small amounts of dust from the primary sampling mirror open the shutter and use photographic type canned air and lens brushes to carefully clean the sam...

Страница 17: ...acetone or isopropanol onto the lens tissue wetting the entire surface of the optic Without lifting the lens tissue drag the tissue across the optic just fast enough so that the solvent evaporates beh...

Страница 18: ...aning or replacement Remove the Leveling Plate to allow access to the bottom of the instrument The Leveling Plate may be removed by carefully prying down on the plate next to each of the leveling feet...

Страница 19: ...instrument The shutter is attached to the front panel with 2 springs Do not let the shutter or springs touch the sampling mirror Figure 3 4 The Mirrors may be removed by removing 3 screws and clamps A...

Страница 20: ...entify it is indicated with an arrow mark on the edge of the mirror Figure 3 5 Before reassembly make sure that all mechanical and optical surfaces are clean and free of dust Reassemble the instrument...

Страница 21: ...o prevent water damage to the measurement head purge air must be provided DO NOT USE DE IONIZED DI WATER DI WATER WILL QUICKLY CAUSE SEVERE CORROSION OF ALL COMPONENTS THAT CONTACT THE WATER Water Pre...

Страница 22: ...1 8 30 9 87 6 35 2 95 3 39 4 102 9 7 6 2 0 30 5 86 9 34 3 93 7 38 1 100 6 8 3 2 2 30 1 86 2 33 6 92 5 37 1 98 7 9 1 2 4 29 8 85 7 33 0 91 5 36 2 97 2 9 8 2 6 29 6 85 3 32 5 90 6 35 5 95 9 10 6 2 8 32...

Страница 23: ...0 561 6 4 000 160 16 0 808 7 4 000 220 22 1 1 8 4 000 290 29 1 4 9 4 000 360 36 1 8 10 4 000 450 45 2 2 11 4 000 540 54 2 7 12 4 000 650 65 3 2 13 4 000 760 76 3 8 14 4 000 880 88 4 4 15 4 000 1 010...

Страница 24: ...ower Energy Diameter 2 4 8 15 30 60 120 250 500 1000 2000 4000 Watts mm 0 03 0 06 0 13 0 25 0 5 1 2 4 8 16 32 64 125 250 500 1000 Joules 4 0 1 1 1 1 2 2 2 3 3 3 4 4 4 4 5 0 1 1 1 2 2 2 2 3 3 3 4 4 4 6...

Страница 25: ...C 2 IBP YAG Laser settings Appendix C Calibration Information...

Отзывы: