Removing and Replacing the sampling optics 

Replacement of the sampling optics involves partial disassembly of the IBP-YAG.  The following 
procedure will allow access to the sampling optics for cleaning or replacement. 
 
Remove the Leveling Plate to allow access to the bottom of the instrument.  The Leveling Plate 

may be removed by carefully prying down on the plate next to each of the leveling feet  The studs 

of the leveling feet will pop out of the pads attached to the plate. 
Remove the screws in the cover and remove the cover by sliding back and up. 

 

Figure 

 

3-2 

 

IBP-YAG Maintenance 

and 

Storage 

3-3 

Содержание IBP-YAG Series

Страница 1: ...W 1000 N Main Logan UT 84321 For Sales Service or Technical Support Phone 435 753 3729 Fax 435 753 5231 Email service ophir spiricon com Email sales ophir spiricon com Copyright 2008 Ophir Spiricon In...

Страница 2: ...manual at any time and without notice All rights to the product and any accompanying operators manuals are reserved While every precaution has been taken in the preparation of this product the publis...

Страница 3: ...__________ 2 1 2 3 Frame Grabber Setup ________________________________________2 2 2 3 1 Refer to your LBA PC Operators Manual ___________________________ 2 2 2 3 2 Starting LBA PC for the first time_...

Страница 4: ...and use procedures We also suggest that you read the software manual to become familiar with the controls and the measurements provided Chapter 1 contains general Information and specifications of the...

Страница 5: ...ndensing Note When storage temperature may drop below 5 C all cooling water must be removed from the unit to prevent damage due to freezing Operating Temperature 10o C to 50o C 50 F to 122 F Operating...

Страница 6: ...ons Weight 12 kg 26 lb Dimensions 24 x 50 x 32 cm 9 6 x 19 5 x 12 6 17 8cm 7 0 32 1cm 12 6 22 8cm 9 0 26 0cm 10 2 48 4cm 19 1 24 3cm 9 6 19 7cm 7 8 45 7cm 18 0 Figure 1 1 Overall Dimensions IBP YAG Ge...

Страница 7: ...to connect the IBP YAG to the LBA 7xx frame grabber card 3 Cooling Water Inlet 4 Cooling Water Outlet These fittings are standard barbed hose connectors for 5 16 ID Hose Refer to appendix A for water...

Страница 8: ...e shutter closed except when the laser is directed into this opening 9 Shutter Release Lift up on the top handle to close the shutter 10 Bubble Level 11 Leveling Adjusters Use the leveling adjusters t...

Страница 9: ...tive eyewear when operating the laser Do not set up the IBP YAG or any optical components at eye level Do not wear reflective jewelry or watches while operating this equipment to reduce the possibilit...

Страница 10: ...nal Cart system If ordered the cart system will be shipped separately and will include separate unpacking and setup instructions 2 2 Setting up the IBP YAG 2 2 1 Frame Grabber and Software Setup The I...

Страница 11: ...this connection remove the plug from the back of the instrument and replace it with an 1 8 NPT hose fitting Connect to a supply of clean dry nitrogen or air at about 5 psi 2 2 2 3 Power and Video conn...

Страница 12: ...dialog box and under Method select the Continuous option Then click OK If there still is no indication of operation verify in the lower status bar that the Video V enunciator is illuminated Green If t...

Страница 13: ...tions or when using chilled cooling water run the purge air for 10 min before starting the cooling water Set fine attenuation knob to the minimum position by turning the knob clockwise until it stops...

Страница 14: ...saturated portions of the displayed beam will turn white when saturated Increase the fine attenuation until beam is just below saturation Center the Beam image in the display by adjusting the levelin...

Страница 15: ...r s shutter or by turning the laser off Never attempt to block the laser beam by closing the IBP YAG shutter The IBP YAG shutter is not designed to handle high energy beams and doing so will damage th...

Страница 16: ...cs 3 2 1 Cleaning the sampling optics To remove small amounts of dust from the primary sampling mirror open the shutter and use photographic type canned air and lens brushes to carefully clean the sam...

Страница 17: ...acetone or isopropanol onto the lens tissue wetting the entire surface of the optic Without lifting the lens tissue drag the tissue across the optic just fast enough so that the solvent evaporates beh...

Страница 18: ...aning or replacement Remove the Leveling Plate to allow access to the bottom of the instrument The Leveling Plate may be removed by carefully prying down on the plate next to each of the leveling feet...

Страница 19: ...instrument The shutter is attached to the front panel with 2 springs Do not let the shutter or springs touch the sampling mirror Figure 3 4 The Mirrors may be removed by removing 3 screws and clamps A...

Страница 20: ...entify it is indicated with an arrow mark on the edge of the mirror Figure 3 5 Before reassembly make sure that all mechanical and optical surfaces are clean and free of dust Reassemble the instrument...

Страница 21: ...o prevent water damage to the measurement head purge air must be provided DO NOT USE DE IONIZED DI WATER DI WATER WILL QUICKLY CAUSE SEVERE CORROSION OF ALL COMPONENTS THAT CONTACT THE WATER Water Pre...

Страница 22: ...1 8 30 9 87 6 35 2 95 3 39 4 102 9 7 6 2 0 30 5 86 9 34 3 93 7 38 1 100 6 8 3 2 2 30 1 86 2 33 6 92 5 37 1 98 7 9 1 2 4 29 8 85 7 33 0 91 5 36 2 97 2 9 8 2 6 29 6 85 3 32 5 90 6 35 5 95 9 10 6 2 8 32...

Страница 23: ...0 561 6 4 000 160 16 0 808 7 4 000 220 22 1 1 8 4 000 290 29 1 4 9 4 000 360 36 1 8 10 4 000 450 45 2 2 11 4 000 540 54 2 7 12 4 000 650 65 3 2 13 4 000 760 76 3 8 14 4 000 880 88 4 4 15 4 000 1 010...

Страница 24: ...ower Energy Diameter 2 4 8 15 30 60 120 250 500 1000 2000 4000 Watts mm 0 03 0 06 0 13 0 25 0 5 1 2 4 8 16 32 64 125 250 500 1000 Joules 4 0 1 1 1 1 2 2 2 3 3 3 4 4 4 4 5 0 1 1 1 2 2 2 2 3 3 3 4 4 4 6...

Страница 25: ...C 2 IBP YAG Laser settings Appendix C Calibration Information...

Отзывы: