HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual 

was furthest out of range for the sensor before the alarm 
cleared (see the Points table for the actual value). 

 

Alarms can only be configured on the thermocouple 
channel; they are not available for the internal 10K 
thermistor (temperature) channel. 

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, let’s say the logger is recording data at a 5-minute 
logging interval and burst logging is configured to log every 10 
seconds when the temperature goes above 2,000°F (the high 
limit) or falls below -200°F (the low limit). This means the logger 
will record data every 5 minutes as long as the temperature 
remains between 2,000°F and -200°F. Once the temperature 
reaches 2,100°F, for example, the logger will switch to the 
faster logging rate and record data every 10 seconds until the 
temperature falls back below the high limit (or 2,000°F in this 
case). At that time, logging then resumes every 5 minutes at 
the normal logging interval. Similarly, if the temperature falls to 
-250°F, for example, then the logger would switch to burst 
logging mode again and record data every 10 seconds. Once the 
temperature rises back to -200°F, the logger will then return to 
normal mode, logging every 5 minutes. To set up burst logging: 

1.

 

Select Burst for Logging Mode in the Launch Logger 
window. If Burst has already been configured for this 
logger, click the Edit button in the Launch Logger window. 
Note that you cannot set up alarms with burst logging. You 
also cannot select the Stop Logging option “Never (wrap 
when full)” if burst logging is configured. 

2.

 

In the Burst Logging window, select the High Limit checkbox 
if you want to set up a condition in which burst logging will 
occur when the sensor reading rises above the high limit 
value. Type in the value or drag the red upper slider. 

3.

 

Select the Low Limit checkbox if you want to set up a 
condition in which burst logging will occur when the sensor 
reading falls below the low limit value. Type in the value or 
drag the blue lower slider. 

 

4.

 

Set the burst logging interval, which must be less than the 
logging interval. Select either a preset burst logging interval 

or select Custom and enter your own interval. Keep in mind 
that the more frequent the burst logging rate, the greater 
the impact on battery life and the shorter the logging 
duration. 

5.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging Mode 
in the Launch Logger window to make additional changes. 

Notes: 

 

Burst logging is available on the thermocouple channel 
only; it is not available for the internal 10K thermistor 
(temperature) channel. 

 

Once the logger is launched, the high and low burst 
logging limits are only checked when the logger's 
LCD screen refreshes once every 15 seconds. Therefore, 
if you set the logging interval to less than 15 seconds and 
the sensor reading falls outside the limits, the burst 
logging will not begin until the next 15-second refresh 
cycle. 

 

The actual values for burst logging limits are set to the 
closest values supported by the logger based on the 
thermocouple type. In addition, burst logging mode can 
begin or end when the sensor reading is within the 
resolution specification for each thermocouple type. This 
means the value that triggers the alarm may differ 
slightly than the value entered. 

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded in “normal mode.” For example, let’s assume 
the logger has a 10-minute logging interval and logged a 
data point at 9:05. Then, the high limit was surpassed 
and burst logging began at 9:06. Burst logging then 
continued until 9:12 when the sensor reading fell back 
below the high limit. Now back in normal mode, the next 
logging interval will be 10 minutes from the last burst 
logging point, or 9:22 in this case. If burst logging had not 
occurred, the next data point would have been at 9:15. 

 

A New Interval event will appear on the plot (if you select 
events for plotting in the Plot Setup window) each time 
the logger enters or exits burst logging mode. 

Statistics 

Statistics is a logging mode in which the logger calculates 
maximum, minimum, average, and standard deviation statistics 
during logging, recording the results at each logging interval 
based on samples taken at a rate you specify. This will result in 
up to four additional series per sensor that record the following 
information at each logging interval: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 
values. 

For example, let’s say the logging interval is set to 5 minutes 
and the sampling interval is set to 30 seconds (with maximum, 
minimum, average, and standard deviation all enabled). Once 
logging begins, the logger will measure and record the actual 
temperature sensor values every 5 minutes. In addition, the 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание HOBO UX100-014M

Страница 1: ...F thermocouple probe accuracy 0 08 C 0 15 F S 50 to 1 720 C 58 to 3 128 F 2 2 C 3 96 F thermocouple probe accuracy 0 08 C 0 15 F B 550 to 1 820 C 1 022 to 3 308 F 2 5 C 4 5 F thermocouple probe accura...

Страница 2: ...button for 1 second to record an internal event see Recording Internal Logger Events or to turn the LCD screen on if the option to turn off the LCD has been enabled see Setting up the Logger Battery T...

Страница 3: ...atistics in HOBOware Press the Alarm Stats button for 1 second to cycle through each of the available statistics and then back to the current temperature or to the alarm value if applicable This is th...

Страница 4: ...g the oldest This option is not available if Burst is selected for the Logging Mode Push Button Logging will end once you press the Start Stop logging button for 3 seconds Note that if you also choose...

Страница 5: ...e course of the deployment is equal to the selected duration If you select Consecutive then the alarm will trip when the time the sensor is continuously out of range is equal to the selected duration...

Страница 6: ...ogger window Click the Edit button next to Logging Mode in the Launch Logger window to make additional changes Notes Burst logging is available on the thermocouple channel only it is not available for...

Страница 7: ...OBOware or connect it to a HOBO U Shuttle U DT 1 firmware version 1 18m030 or higher and then offload the data files from the U Shuttle to HOBOware Refer to the HOBOware Help for details Recording Int...

Страница 8: ...r to the computer click the Readout button on the toolbar and save the data as prompted Replace the battery before launching the logger again To replace the battery 1 Holding the logger with the LCD s...

Отзывы: