Onset HOBO UX100-014M Скачать руководство пользователя страница 3

 

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual 

USB Port:

 Use this port (not visible in the diagram) to connect 

the logger to the computer or the HOBO U-Shuttle via USB 
cable (see 

Setting up the Logger 

and 

Reading Out the Logger

). 

LCD Screen:

 This logger is equipped with an LCD screen that 

displays details about the current status. This example shows all 
symbols illuminated on the LCD screen followed by definitions 
of each symbol in the table below. 

 

LCD Symbol 

Description 

 

The logger is waiting to be launched. Press and 
hold the Start/Stop button for 3 seconds to launch 
the logger. 

 

The logger has been launched with push button 
stop enabled; press and hold the Start/Stop 
button for 3 seconds to stop the logger. 

Note

: If 

you also launched the logger with a push button 
start, this symbol will not appear on the display for 
30 seconds. 

 

The battery indicator shows the approximate 
battery power remaining. 

 

The logger has been configured to stop logging 
when memory fills. The memory bar indicates the 
approximate space remaining in the logger to 
record data. When first launched, all five segments 
in the bar will be empty. In this example, the 
logger memory is almost full (only one segment in 
the memory bar is empty). 

 

The logger has been configured to never stop 
logging (wrapping). The logger will continue 
recording data indefinitely, with newest data 
overwriting the oldest data. When first launched, 
all five segments in the memory bar will be empty. 
In this example, the memory is full (all five 
segments are filled in) and new data is now 
overwriting the oldest data. This will continue until 
the logger is stopped or the battery runs out. 

 

The logger is currently logging. 

 

A sensor reading is above or below the high or low 
alarm that you configured. Press and release the 
Alarm/Stats button until the “alm” symbol 
(described below) is displayed on the screen. This 
symbol at left will clear depending on how alarms 
were configured in HOBOware. If the alarm was 
configured to clear when the logger is relaunched, 
this symbol will remain on the LCD. Otherwise, it 
will clear when the sensor reading is back within 
the alarm limits or by pressing the Alarm/Stats 
button for 3 seconds. 

 

An alarm is ready to be cleared. This will only 
appear if “Cleared with button press” was selected 
in the HOBOware alarm settings. Press the 
Alarm/Stats button for 3 seconds to clear the 
alarm. 

 

 

These symbols show the maximum, minimum, 
average, and standard deviation values most 
recently calculated for the logger (if the logging 

LCD Symbol 

Description 

 

 

mode has been set to Statistics in HOBOware). 
Press the Alarm/Stats button for 1 second to cycle 
through each of the available statistics and then 
back to the current temperature (or to the alarm 
value if applicable). 

 

This is the sensor reading that tripped the alarm. 
Press the Alarm/Stats button to view this reading. 
Press the Alarm/Stats button again to cycle 
through any statistics (defined above) and 
ultimately back to the current temperature. 

 

This is an example of a temperature reading. 
Temperature units are determined by the settings 
in HOBOware. To switch between Celsius and 
Fahrenheit, change the Display Preferences in 
HOBOware before launching the logger. 

 

The logger has been configured to start logging on 
a particular date/time. The display will count down 
in days, hours, minutes, and seconds until logging 
begins. In this example, 5 minutes and 38 seconds 
remain until logging will begin. 

 

The launch settings are being loaded onto the 
logger from HOBOware. Do not disconnect the 
USB cable during this process. 

 

An error occurred while loading the launch 
configurations onto the logger from HOBOware. 
Make sure the USB cable is connected to both the 
logger and the computer and try launching again. 

 

The logger has been stopped with HOBOware or 
because the memory is full. 

Notes: 

 

You can disable the LCD screen when logging. Select 
“Turn LCD off” when setting up the logger as described in 
the next section. When this option is enabled, you can 
still temporarily view the LCD screen by pushing the 
Start/Stop button for 1 second. The LCD will then remain 
on for 10 minutes. 

 

The LCD screen refreshes every 15 seconds regardless of 
the logging interval selected in HOBOware. If you choose 
a logging interval less than 15 seconds, the data will be 
recorded at the faster interval, but the sensor readings 
will only be updated on the screen every 15 seconds. 

 

When the logger has stopped logging, the LCD screen will 
remain on until the logger is offloaded to a computer or 
HOBO U-Shuttle (unless launched with the “Turn LCD off” 
option). Once the logger has been offloaded and 
disconnected from the computer, the LCD will turn off 
automatically after 2 hours. The LCD will turn back on the 
next time the logger is connected to the computer. 

Setting up the Logger 

Use HOBOware to set up the logger, including setting alarms, 
selecting the options to start and stop logging, and choosing a 
logging mode. 

1.

 

Connect the logger and open the Launch Logger window

To connect the logger to a computer, use the USB cable 
provided. Click the Launch icon on the HOBOware toolbar 
or select Launch from the Device menu. 

 

Important:

 USB 2.0 specifications do not guarantee 

operation outside the range of 0°C (32°F) to 50°C (122°F). 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание HOBO UX100-014M

Страница 1: ...F thermocouple probe accuracy 0 08 C 0 15 F S 50 to 1 720 C 58 to 3 128 F 2 2 C 3 96 F thermocouple probe accuracy 0 08 C 0 15 F B 550 to 1 820 C 1 022 to 3 308 F 2 5 C 4 5 F thermocouple probe accura...

Страница 2: ...button for 1 second to record an internal event see Recording Internal Logger Events or to turn the LCD screen on if the option to turn off the LCD has been enabled see Setting up the Logger Battery T...

Страница 3: ...atistics in HOBOware Press the Alarm Stats button for 1 second to cycle through each of the available statistics and then back to the current temperature or to the alarm value if applicable This is th...

Страница 4: ...g the oldest This option is not available if Burst is selected for the Logging Mode Push Button Logging will end once you press the Start Stop logging button for 3 seconds Note that if you also choose...

Страница 5: ...e course of the deployment is equal to the selected duration If you select Consecutive then the alarm will trip when the time the sensor is continuously out of range is equal to the selected duration...

Страница 6: ...ogger window Click the Edit button next to Logging Mode in the Launch Logger window to make additional changes Notes Burst logging is available on the thermocouple channel only it is not available for...

Страница 7: ...OBOware or connect it to a HOBO U Shuttle U DT 1 firmware version 1 18m030 or higher and then offload the data files from the U Shuttle to HOBOware Refer to the HOBOware Help for details Recording Int...

Страница 8: ...r to the computer click the Readout button on the toolbar and save the data as prompted Replace the battery before launching the logger again To replace the battery 1 Holding the logger with the LCD s...

Отзывы: