Connecting the Thermocouple Sensor 

The logger can accept any standard thermocouple sensor of the 
appropriate type with standard thermocouple subminiature 
connectors. 

ANSI Standard Thermocouple Color Coding 

Type-J Iron-Constantan 

Black 

Type-K Chromel-Alumel 

Yellow 

Type-S Platinum-Rhodium 

Green 

Type-T Copper-Constantan 

Blue 

Type-E Chromel-Constantan 

Purple 

Type-R Platinum-Rhodium 

Green 

Type-B Platinum-Rhodium 

Gray 

Type-N Nicrosil-Nisil  Orange 

 

Refer to http://www.omega.com/techref/colorcodes.html for 
additional details on color coding. 

If the thermocouple has stripped-wire ends, male subminiature 
connectors with screw terminals are available. You must use 
the matching type of subminiature connector for the 
thermocouple you are using. 

Note that the subminiature connectors have one pin (negative 
terminal) that is wider than the other (positive terminal). Make 
sure the plug is inserted into the jack correctly as shown 

 

Setting up Alarms 

You can set an alarm to trip on the logger when a sensor 
reading rises above or falls below a specified value. To set an 
alarm: 

1.

 

Click the Alarms button from the Launch Logger window. If 
the Alarms button is disabled, make sure the Logging Mode 
is not set to Burst. (Alarms can only be configured if the 
logger is in Normal or Statistics mode.) 

2.

 

Select the High Alarm checkbox if you want an alarm to trip 
when the temperature rises above the high alarm value. 
Type the reading next to the High Alarm checkbox or drag 
the red upper slider in the Configure Alarms window. 

3.

 

Select the Low Alarm checkbox if you want an alarm to trip 
when the temperature falls below the low alarm value. 
Type the reading next to the Low Alarm checkbox or drag 
the blue lower slider. 

4.

 

Set the duration before an alarm is tripped. 

5.

 

Select either Cumulative or Consecutive. If you select 
Cumulative, then the alarm will trip when the time the 
sensor is out of range over the course of the deployment is 
equal to the selected duration. If you select Consecutive, 

then the alarm will trip when the time the sensor is 
continuously out of range is equal to the selected duration. 
For example, the high alarm is set to 300°F and the duration 
is set to 30 minutes. If Cumulative is selected, then an alarm 
will trip once a sensor reading has been at or above 300°F 
for a total of 30 minutes since the logger was configured; 
specifically, this could be 15 minutes above 300°F in the 
morning and then 15 minutes above 300°F again in the 
afternoon. If Consecutive is selected, then an alarm will trip 
only if all sensor readings are 300°F or above for a 
continuous 30-minute period. 

 

6.

 

Choose how long the logger should maintain the sensor 
alarm once it has tripped. Select “Host has relaunched 
logger” if you want the alarm to remain visible on the LCD 
until the next time you relaunch the logger. Select “Sensor 
reading within limits” if you want the alarm to clear once 
the sensor reading returns to the normal range between 
the high and low alarm limits. Select “Cleared with button 
press” if you want the alarm to remain on until you press 
the Alarm/Stats button on the logger. 

7.

 

Click OK to save alarm settings. 

Notes: 

 

Once the logger is launched, alarms will trip as 
determined by these settings. Logger alarms will display 
on the LCD screen. Note that the alarm limits are only 
checked when the logger’s LCD screen refreshes every 15 
seconds. 

 

The actual values for the high and low alarm limits are set 
to the closest values supported by the logger based on 
the thermocouple type. In addition, alarms can trip or 
clear when the sensor reading is within the resolution 
specification for each thermocouple type. This means the 
value that triggers the alarm may differ slightly than the 
value entered.  

 

When you read out the logger, high and low alarm levels 
will be displayed on the plot along with “Chan 1 Alarm 
Tripped” and “Chan 1 Alarm Cleared” events showing 
when the temperature alarm tripped and cleared. The 
“Chan 1 Alarm Cleared” event contains the value that 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание HOBO UX100-014M

Страница 1: ...F thermocouple probe accuracy 0 08 C 0 15 F S 50 to 1 720 C 58 to 3 128 F 2 2 C 3 96 F thermocouple probe accuracy 0 08 C 0 15 F B 550 to 1 820 C 1 022 to 3 308 F 2 5 C 4 5 F thermocouple probe accura...

Страница 2: ...button for 1 second to record an internal event see Recording Internal Logger Events or to turn the LCD screen on if the option to turn off the LCD has been enabled see Setting up the Logger Battery T...

Страница 3: ...atistics in HOBOware Press the Alarm Stats button for 1 second to cycle through each of the available statistics and then back to the current temperature or to the alarm value if applicable This is th...

Страница 4: ...g the oldest This option is not available if Burst is selected for the Logging Mode Push Button Logging will end once you press the Start Stop logging button for 3 seconds Note that if you also choose...

Страница 5: ...e course of the deployment is equal to the selected duration If you select Consecutive then the alarm will trip when the time the sensor is continuously out of range is equal to the selected duration...

Страница 6: ...ogger window Click the Edit button next to Logging Mode in the Launch Logger window to make additional changes Notes Burst logging is available on the thermocouple channel only it is not available for...

Страница 7: ...OBOware or connect it to a HOBO U Shuttle U DT 1 firmware version 1 18m030 or higher and then offload the data files from the U Shuttle to HOBOware Refer to the HOBOware Help for details Recording Int...

Страница 8: ...r to the computer click the Readout button on the toolbar and save the data as prompted Replace the battery before launching the logger again To replace the battery 1 Holding the logger with the LCD s...

Отзывы: