![Onset HOBO UX100-014M Скачать руководство пользователя страница 1](http://html1.mh-extra.com/html/onset/hobo-ux100-014m/hobo-ux100-014m_manual_1638493001.webp)
HOBO® Single Channel Thermocouple
Data Logger (UX100-014M) Manual
16211-G
The HOBO Single Channel Thermocouple data logger records temperature in indoor
environments with J, K, T, E, R, S, B, or N type thermocouple sensors (sold separately). The
logger also has an internal 10K thermistor for cold-junction compensation of the thermocouple
output. The thermocouple channel has a 20-bit resolution while the internal 10K thermistor has
12-bit resolution. Using HOBOware®, you can easily configure the logger alarm to trip for
specific high or low temperatures. Or, you can set up burst logging in which the logger records
data at a different interval during certain conditions. The logger can also calculate minimum,
maximum, average, and standard deviation statistics. This compact data logger has a built-in
LCD screen to monitor temperature, logging status, battery use, and memory consumption in
between readouts.
Specifications
Thermocouple
Type Range
Accuracy
Resolution
J
-210° to 760°C
(-346° to 1,400°F)
±0.6°C (±1.08°F) ± thermocouple
probe accuracy
0.03°C (0.06°F)
K
-260° to 1,370°C
(-436° to 2,498°F)
±0.7°C (±1.26°F) ± thermocouple
probe accuracy
0.04°C (0.07°F)
T
-260° to 400°C
(-436° to 752°F)
±0.6°C (±1.08°F) ± thermocouple
probe accuracy
0.02°C (0.03°F)
E
-260° to 950°C
(-436° to 1,742°F)
±0.6°C (±1.08°F) ± thermocouple
probe accuracy
0.03°C at (0.05°F)
R
-50° to 1,550°C
(-58° to 2,822°F)
±2.2°C (±3.96°F) ± thermocouple
probe accuracy
0.08°C (0.15°F)
S
-50° to 1,720°C
(-58° to 3,128°F)
±2.2°C (±3.96°F) ± thermocouple
probe accuracy
0.08°C (0.15°F)
B
550° to 1,820°C
(1,022° to 3,308°F)
±2.5°C (±4.5°F) ± thermocouple
probe accuracy
0.1°C (0.18F)
N
-260° to 1,300°C
(-436° to 2,372°F)
±1.0°C (±1.8°F) ± thermocouple
probe accuracy
0.06°C (0.11°F)
Internal 10K Thermistor (Temperature)
Range
-20° to 70°C (-4° to 158°F)
Accuracy
±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
Resolution
0.024°C at 25°C (0.04°F at 77°F); see Plot A
Drift
<0.1°C (0.18°F) per year
Logger
Logger Operating Range
Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing);
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification
Logging Rate
1 second to 18 hours, 12 minutes, 15 seconds
Logging Modes
Fixed interval (normal), burst, or statistics
Memory Modes
Wrap when full or stop when full
Start Modes
Immediate, push button, date & time, or next interval
Stop Modes
When memory full, push button, or date & time
Restart Mode
Push button
Time Accuracy
±1 minute per month at 25°C (77°F), see Plot B
Power Source
One 3V CR2032 lithium battery and USB cable
Battery Life
1 year, typical with logging rate of 1 minute and sampling interval of
15 seconds or greater
Memory
512 KB (208,076 measurements, maximum)
HOBO Single Channel
Thermocouple
Data Logger
UX100-014M
Included Items:
•
Command™ strip
•
Double-sided tape
•
Hook & loop strap
Required Items:
•
HOBOware 3.4 or later
•
USB cable (included with
software)
•
Thermocouple sensor type
J, K, T, E, R, S, B, or N
www.
GlobalTestSupply
.com
Find Quality Products Online at: