Onset HOBO UX100-014M Скачать руководство пользователя страница 1

HOBO® Single Channel Thermocouple  
Data Logger (UX100-014M) Manual 

16211-G 

The HOBO Single Channel Thermocouple data logger records temperature in indoor 
environments with J, K, T, E, R, S, B, or N type thermocouple sensors (sold separately). The 
logger also has an internal 10K thermistor for cold-junction compensation of the thermocouple 
output. The thermocouple channel has a 20-bit resolution while the internal 10K thermistor has 
12-bit resolution. Using HOBOware®, you can easily configure the logger alarm to trip for 
specific high or low temperatures. Or, you can set up burst logging in which the logger records 
data at a different interval during certain conditions. The logger can also calculate minimum, 
maximum, average, and standard deviation statistics. This compact data logger has a built-in 
LCD screen to monitor temperature, logging status, battery use, and memory consumption in 
between readouts. 

Specifications 

Thermocouple 

 

Type Range 

Accuracy 

Resolution 

-210° to 760°C  
(-346° to 1,400°F) 

±0.6°C (±1.08°F) ± thermocouple 
probe accuracy 

0.03°C (0.06°F) 

-260° to 1,370°C  
(-436° to 2,498°F) 

±0.7°C (±1.26°F) ± thermocouple 
probe accuracy 

0.04°C (0.07°F) 

-260° to 400°C  
(-436° to 752°F) 

±0.6°C (±1.08°F) ± thermocouple 
probe accuracy 

0.02°C (0.03°F) 

-260° to 950°C  
(-436° to 1,742°F) 

±0.6°C (±1.08°F) ± thermocouple 
probe accuracy 

0.03°C at (0.05°F) 

-50° to 1,550°C  
(-58° to 2,822°F) 

±2.2°C (±3.96°F) ± thermocouple 
probe accuracy 

0.08°C (0.15°F) 

-50° to 1,720°C  
(-58° to 3,128°F) 

±2.2°C (±3.96°F) ± thermocouple 
probe accuracy 

0.08°C (0.15°F) 

550° to 1,820°C  
(1,022° to 3,308°F) 

±2.5°C (±4.5°F) ± thermocouple 
probe accuracy 

0.1°C (0.18F) 

-260° to 1,300°C  
(-436° to 2,372°F) 

±1.0°C (±1.8°F) ± thermocouple 
probe accuracy 

0.06°C (0.11°F) 

Internal 10K Thermistor (Temperature) 

 

Range 

-20° to 70°C (-4° to 158°F) 

Accuracy 

±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A 

Resolution 

0.024°C at 25°C (0.04°F at 77°F); see Plot A 

Drift 

<0.1°C (0.18°F) per year 

Logger 

 

Logger Operating Range 

Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing); 
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification 

Logging Rate 

1 second to 18 hours, 12 minutes, 15 seconds 

Logging Modes 

Fixed interval (normal), burst, or statistics 

Memory Modes 

Wrap when full or stop when full 

Start Modes 

Immediate, push button, date & time, or next interval 

Stop Modes 

When memory full, push button, or date & time 

Restart Mode 

Push button 

Time Accuracy 

±1 minute per month at 25°C (77°F), see Plot B 

Power Source 

One 3V CR2032 lithium battery and USB cable 

Battery Life 

1 year, typical with logging rate of 1 minute and sampling interval of 
15 seconds or greater 

Memory 

512 KB (208,076 measurements, maximum) 

 

HOBO Single Channel 
Thermocouple  
Data Logger 

UX100-014M 

Included Items: 

 

Command™ strip 

 

Double-sided tape 

 

Hook & loop strap 

Required Items: 

 

HOBOware 3.4 or later 

 

USB cable (included with 
software) 

 

Thermocouple sensor type 
J, K, T, E, R, S, B, or N 

 

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание HOBO UX100-014M

Страница 1: ...F thermocouple probe accuracy 0 08 C 0 15 F S 50 to 1 720 C 58 to 3 128 F 2 2 C 3 96 F thermocouple probe accuracy 0 08 C 0 15 F B 550 to 1 820 C 1 022 to 3 308 F 2 5 C 4 5 F thermocouple probe accura...

Страница 2: ...button for 1 second to record an internal event see Recording Internal Logger Events or to turn the LCD screen on if the option to turn off the LCD has been enabled see Setting up the Logger Battery T...

Страница 3: ...atistics in HOBOware Press the Alarm Stats button for 1 second to cycle through each of the available statistics and then back to the current temperature or to the alarm value if applicable This is th...

Страница 4: ...g the oldest This option is not available if Burst is selected for the Logging Mode Push Button Logging will end once you press the Start Stop logging button for 3 seconds Note that if you also choose...

Страница 5: ...e course of the deployment is equal to the selected duration If you select Consecutive then the alarm will trip when the time the sensor is continuously out of range is equal to the selected duration...

Страница 6: ...ogger window Click the Edit button next to Logging Mode in the Launch Logger window to make additional changes Notes Burst logging is available on the thermocouple channel only it is not available for...

Страница 7: ...OBOware or connect it to a HOBO U Shuttle U DT 1 firmware version 1 18m030 or higher and then offload the data files from the U Shuttle to HOBOware Refer to the HOBOware Help for details Recording Int...

Страница 8: ...r to the computer click the Readout button on the toolbar and save the data as prompted Replace the battery before launching the logger again To replace the battery 1 Holding the logger with the LCD s...