FT-3100 TRANSMITTER PROGRAMMING GUIDE
ONICON Incorporated 727.447.6140
Page 27
onicon.com
When enabled, this function allows the autozero calibration function to be
remotely activated via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode.
When enabled, this function allows the pulse input to be used to remotely
change the flow range to full scale flow rate 2 [FS2 = Gal/m XXX.X] in SCALES
menu (menu 3).
Note: this function cannot be used when the autorange [Autorange = OFF] function
in MEASURE menu (menu 4) is enabled.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode.
Activate Autozero Calibration
via Pulse Input
[Calibration = OFF]
Change Flow Range via
Pulse Input
[Range change = OFF]