NOTE:
If the actions listed below do not resolve the error, call LifeScan Customer Services at
1
(
800
)
227
-
8862
.
ERROR
POTENTIAL PROBLEM
WHAT TO DO
1
.
Repeat the test with a new test strip and:
◆
Check the test strip expiration date. Do not use
test strips beyond the expiration or discard
date.
The meter detected
a problem with the
test strip.
Professional Use:
◆
Test the sample with an alternate method.
While awaiting results:
Compare the color of the test strip confirmation dot
to the Color Chart on the test strip bottle label. If the
glucose level is very low (
50
mg/dL) or very high
(
350
mg/dL), take appropriate precautions immediately.
Before retesting verify that the sample hematocrit is
within range.
◆
If these actions do not resolve the error, call LifeScan
Customer Services at
1 800 227
–
8862
.
The meter detected
a problem with the
color of the test
strip.
Er 1
continued
50
TROUBLESHOOTING
DCO_05074503A_OT_SS_OB_US 6/19/03 11:59 AM Page 50