![OneTouch ONETOUCH SURESTEP (WITH DATA PORT) Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/onetouch/onetouch-surestep-with-data-port/onetouch-surestep-with-data-port_owners-booklet_1635832027.webp)
17
CHECKING
NOTE:
If opening the test strip bottle for
the first time, write the discard date
(
4
months from today) on the bottle
label and perform a control solution test
following the instructions below.
4
.
Shake the control solution bottle. Apply
1
drop of control solution to the center of
the pink test square.
◆
The first time you open the control
solution bottle, write the discard date
on the label in the space provided. The
discard date for control solution is
3
months from the date you first open
the bottle.
◆
Check expiration date and discard
date. Do not use the control solution
after these dates.
◆
If the entire white pad surrounding the
pink test square is soaked, you have
applied too much Glucose Control
Solution.
CAUTION:
Applying too much control
solution may cause out–of–range results
and may make the meter dirty.