41
MAINTENANCE
General Care
◆
Keep your meter dry and avoid exposing
it to extremes in temperature. Do not
store your meter or test strips in your car
or in the refrigerator or freezer.
◆
Dropping the meter can damage the
electronics. If you drop your meter, do a
control solution test (see
Checking the
System
) and a display segment check
(see
Display Symbols
) before using the
meter to test your blood.
◆
Do not take the meter apart. Sensitive
parts could be damaged causing inaccu-
rate results. Taking the meter apart will
also void your warranty.
Cleaning the Meter
C A U T I O N :
◆
Use water to clean the meter.
◆
Do not get water inside the meter case.
◆
Never immerse the meter or hold it
under running water.
◆
Do not
use alcohol, glass cleaners, or
any other cleansers to clean the test
strip holder or lens area.
N OT E :
For Healthcare Professionals or
caregivers who assist multiple users, to
disinfect
t
he meter, cleanse the meter as
described below using a
10
% bleach solu-
tion (
9
parts water,
1
part bleach) and
then clean again with water to remove
any remaining bleach. Thoroughly dry
the meter. Failure to remove the excess
bleach may result in an error message
(
,
) or an inaccurately high result.
Caring for Your Meter
DCO_05074503A_OT_SS_OB_US 6/19/03 11:59 AM Page 41
Содержание ONETOUCH SURESTEP (WITH DATA PORT)
Страница 2: ...DCO_05074503A_OT_SS_OB_US 6 19 03 11 59 AM Page D...
Страница 10: ...DCO_05074503A_OT_SS_OB_US 6 19 03 11 59 AM Page viii...
Страница 11: ...DCO_05074503A_OT_SS_OB_US 6 19 03 11 59 AM Page 1...
Страница 86: ...DCO_05074503A_OT_SS_OB_US 6 19 03 11 59 AM Page 76...
Страница 87: ...DCO_05074503A_OT_SS_OB_US 6 19 03 11 59 AM Page 77...