Troubleshooting

Trouble

Possible cause

Corrective action

The nebulisation rate is 
extremely low.

Batteries are low (the orange 
battery low indicator fl ashes).

Replace the batteries with new 
ones. Refer to page 9.

The mesh is stained or clogged.

1. Clean the mesh by boiling. Refer  

to page 16.

2. Replace the mesh cap with a new 

one. Refer to page 11.

The green power 
indicator does not turn 
on and the unit does not 
nebulise.

The batteries are low.

Replace the batteries with new 
ones. Refer to page 9.

The capacity of rechargeable 
batteries is low.

Charge the batteries with a 
commercially available recharger.

The batteries are inserted in the 
wrong direction.

Re-insert the batteries in the correct 
direction. Refer to page 9.

The AC adapter is not correctly 
connected to the main unit.

Remove the AC adapter from the 
main unit and connect it correctly.

The medication container is not 
installed correctly to the main unit.

Remove the medication container 
from the main unit and install 
correctly. Refer to page 12.

The green power 
indicator turns on 
but the unit does not 
nebulise.

The vibrator is not immersed in the 
medication.

Tilt the main unit in order to immerse 
the vibrator in the medication.

The medication container is not 
fi lled with medication.

Fill with medication. Refer to page 
12.

The mesh is broken.

Replace with a new one. Refer to 
page 11.

Содержание Micro A-I-R U22

Страница 1: ...Instruction manual English Deutsch Nederlands U22 IM NE U22 E 04 08 2010 1665580 0C NE U22 E_C_en de nl_M11_100824 pdf...

Страница 2: ...shown on the packaging or in this instruction manual They will be pleased to assist you OMRON HEALTHCARE EUROPE B V Intended use 3 Safety precautions 4 Features and benefits 5 Contents of the OMRON M...

Страница 3: ...nscious or are not breathing spontaneously Environment This product is intended for use in a medical facility such as Hospital clinic and doctor s office in a room of general household and in open air...

Страница 4: ...he main unit and AC adapter with water do not immerse them in water Do not drop medication on the main unit or on the AC adapter If you drop medication immediately wipe it off Do not apply strong shoc...

Страница 5: ...just two penlite batteries your medication is always within reach wherever you go whatever you do Inhalation is possible at various angles The device enables inhalation at various angles with stable n...

Страница 6: ...Unit Medication Container Mouthpiece Mask Adapter Mesh Cap with Mesh Instruction Manual Warranty Card Optional item not included contact your OMRON dealer AC Adapter Child Mask PVC Adult Mask PVC Sof...

Страница 7: ...Stabilizes nebulisation 6 Medication Port Medication can be filled here by opening the mesh cap 7 Vibrator The tip of this vibrator oscillates at high frequency and pushes the medication through the p...

Страница 8: ...ith the vibrator 13 Battery Low Indicator An orange light blinks when the batteries are low 14 ON OFF button 15 Electrodes Connects the vibrator with the main unit 16 Power indicator A green light tur...

Страница 9: ...acity and condition of the batteries the unit can operate up to 8 days on a set of batteries based on usage of 30 minutes a day When the battery low indicator orange flashes the batteries should be re...

Страница 10: ...nnector Press firmly until it clicks Plug in the power plug of the AC adapter to the electrical outlet To remove the AC adapter connector from the main unit remove the power plug from the electrical o...

Страница 11: ...he medication container from the main unit To do so push on the release buttons on both sides This will disengage the latch mechanism Medication container release Buttons Lift up the Mesh Cap by placi...

Страница 12: ...capacity is 7 ml 5 Install the mesh cap Do not touch the mesh The mesh cap seals the medication bottle Confirm if the cap is securely installed 6 Install the medication container on the main unit Clic...

Страница 13: ...main unit 8 Install the mask or mouthpiece on the mask adapter The device is now ready for use Refer to the next section on how to select the nebulisation mode and how to inhale How to assemble the un...

Страница 14: ...bulisation mode If you hold the ON OFF I button for approx 2 seconds or more the unit nebulises in the manual nebulisation mode The unit only nebulises while you keep on pressing the I button and stop...

Страница 15: ...ach deep in to the bronchial tubes Hold your breath briefly then breathe out slowly while removing the mouthpiece from your mouth Do not breathe too quickly Pause when you feel like having a rest You...

Страница 16: ...d mask 1 Remove the mask adapter from the main unit 2 Remove the medication container from the main unit 3 Remove the mesh cap from the medication container discard the medication and open the medicat...

Страница 17: ...the mesh with a cotton swab or with a pin 6 Shake off excessive water from the parts and let them air dry on a clean lint free paper towel 7 Clean the main unit and it s electrodes by wiping it off w...

Страница 18: ...nutes of boiling shake off excessive water from the parts and let them air dry on a clean lint free paper towel 2 Disinfection by using alcohol All parts can be disinfected by using alcohol Note Alcoh...

Страница 19: ...lorhexidine Amphoteric surfactant Disinfecting ethanol Milton 0 1 Osvan 0 1 Hibitane 0 5 Tego 0 2 Mouthpiece SEBS Mask Adapter PC Medication Container Body PSF Horn Titanium Hinges Titanium Packing Si...

Страница 20: ...ay clog the tiny holes in the mesh cap Leave the mesh cap in a white vinegar solution 1 part vinegar diluted with 3 parts of water for 1 hour In addition the mesh cap may be kept overnight in a cup of...

Страница 21: ...mmercially available recharger The batteries are inserted in the wrong direction Re insert the batteries in the correct direction Refer to page 9 The AC adapter is not correctly connected to the main...

Страница 22: ...ting Trouble Possible cause Corrective action The green power indicator turns on but the unit does not nebulise Medication or water has collected on the electrodes on the main unit Remove the medicati...

Страница 23: ...tainer Mesh Cap Mask Adapter Mouthpiece 2 Batteries AA LR6 Child Mask PVC Adult Mask PVC Soft Pouch Instruction Manual Warranty Card Notes Subject to technical modification without prior notice This O...

Страница 24: ...icle size distribution and aerosol output may vary by combination of product medication and ambient conditions such as temperature humidity and atmospheric pressure Performance may vary with drugs suc...

Страница 25: ...lectromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601 1 2 2007 standard...

Страница 26: ...types of wastes and recycle it responsibly to promote the sustainable reuse of material resources Household users should contact either the retailer where they purchased this product or their local go...

Страница 27: ...27 Optional parts and accessories Adult Mask PVC A040002 Child Mask PVC A040003 Main Unit Cover P040012 Mesh Cap P040014 AC Adapter 4125 0380 ADC A040004 Mask Adapter P040015 Mouthpiece P040016...

Страница 28: ...are com Production facility OMRON MATSUSAKA Co Ltd 1855 370 Kubo cho Matsusaka city Mie prefecture 515 8503 JAPAN Subsidiary OMRON HEALTHCARE UK LIMITED Opal Drive Fox Milne Milton Keynes MK15 0DG U K...

Отзывы: