K-50
Temperature Meter
K3MA-L
Temperature Input Shift
Input shift equivalent to the setting value supported for all points within
the sensor measurement range.
Parameter Initialization
This function returns all of the parameters to their initial values.
Use this to reset the K3MA-L after returning it to its factory-set condi-
tion.
Average Processing
Average processing stabilizes displayed values to minimize flicker by
averaging the fluctuating input signals. Average processing can be
performed for the measurement values in either of four steps (OFF, 2
times, 4 times, or 8 times).
This is useful for ignoring rapid fluctuations, e.g., eliminating spike
noise.
Hysteresis (Comparative Output
Models Only)
The hysteresis of comparative outputs can be set to prevent chatter-
ing in the output when the measurement value fluctuates finely near
the OUT value.
Changing the Display Color
The color of the value displayed can be set to either red or green. For
comparative output models, the display color can be set to change
from green to red, or from red to green, according to the status of the
comparison criterion.
Display Auto-return Time
This function automatically returns the display to the operation level’s
current value if no keys are pressed for a preset time (called the dis-
play auto-return time).
Move-to-Protect-Level Time
The time required to shift to the protect level can be set as desired.
MAX/MIN Display
The maximum and minimum measurement (display) values from the
time the power is turned ON until the current time can be stored and
displayed. This is useful, for example, when measuring the maximum
value.
Parameter
Setting
ins
-1999 to 9999
Parameter
Setting
Meaning
init
off
---
on
Initializes all parame-
ters.
Temperature
Before shift
After shift
Temperature input
shift value
Input
Upper limit (high acting)
Set value
Measurement value
Output
ON
OFF
Hysteresis
192.0
258.9
OUT1 value
Green
Red
Current value
MAX value
MIN value
MAX/MIN
MAX/MIN
MAX/MIN
Max
Min
MAX/MIN
LEVEL
MODE
SHIFT
UP
MAX/MIN
LEVEL
MODE
SHIFT
UP
MAX/MIN
LEVEL
MODE
SHIFT
UP