9.1 Specification of HGM-111 meter
◆
Measuring range: 20~600mg/dL
◆
Calibration: Plasma-calibration
◆
Measuring unit: mg/dL
◆
Measuring time: 5 seconds
◆
Sample type: Capillary whole blood samples drawn from the fingertips
or palm
◆
Blood Sample size: 1μL
◆
Operating temperature: (10 C~40 C)/(50 ºF ~104ºF)
◆
Relative operating humidity: Below 90%RH
◆
Memory capacity: Up to (512) test results (together with date, time
and unit) include average results for the last 7, 14 and 30 days
◆
Dimension: 93 mm x 52.3 mm x 25.3 mm (L×W×H)
◆
Weight: 70 g with batteries
◆
Power Supply: two CR2032 batteries
◆
Transmission: users can upload data to PC via transmission function.
◆
Storage temperature: -20 C~50 C (-4 F~122 F)
◆
Please protect the HGM-111 from impact
◆
Please do not disassemble the HGM-111
Chapter 9 Specifications
Samples
Mean glucose concentration measured CV
Sample 1
38.94 mg/dL
84.14 mg/dL
121.83mg/dL
178.24mg/dL
177.52mg/dL (9.85 mmol/L)
334.75mg/dL
177.52mg/dL (9.85 mmol/L)
43.98 mg/dL
136.62 mg/dL
338.71 mg/dL
Sample 2
4.83%
Sample 3
4.87%
Sample 4
4.16%
Sample 5
3.08%
Control solution 1
Control solution 2
4.90%
Control solution 3
4.46%
SD= 2.81 mg/dL
4.01 mg/dL
Within Run Precision:
Between Day Precision:
31
BLOOD GLUCOSE MONITORING SYSTEM-HGM-111