3.1 Preparation before glucose testing
Please prepare materials you need as listed below and read this
instruction manual or test strip manual to ensure the correct measuring
process and results.
□ HGM-111 Blood Glucose Meter
□ HGM-STP1A Blood Glucose Test Strip
□ Antiseptic wipes or alcohol wipes
□ Lancing Device
□ Lancet
3.2 Fingertip blood measurement
Structure:
Chapter 3 Finger Blood Glucose Testing
A. Install the Lancet
1. Unscrew the adjustable tip of the lancing device. Install a new lancet
into the carrier correctly and make sure it is fully seated. Then remove
the protective cap and replace the adjustable tip.
2. Adjust the puncture depth by the selection dial (Normally at the 2nd
level), twist to a suitable number. (Smaller numbers are for shallower
punctures and larger numbers are for deeper puncture)
2
1
3
4
Ejector
Barrel
Trigger
Carrier
Adjustable tip
AST Cap
14
BLOOD GLUCOSE MONITORING SYSTEM-HGM-111