4.1 HGM-111 system testing
In order to provide you with an accurate test result, we suggest running
a control test to check if the meter and test strips are working properly.
When the control result is within the range on the test strip container, the
test strips and meter are working properly.
◆
Important:
□ Use Only the OMRON control solution with the HGM-111 meter and
HGM-STP1A strips.
□ Mark the newly opened bottle of control solution with the date opened;
discard any unused control solution which is over three months after
opening.
□ Store the bottle at 4
o
C~30 C (39 F~86 F) temperature. Do not
refrigerate it.
□ Shake the bottle of the control solution before you use it. Please
replace new control solution if control solution deteriorates or
coagulates.
* Before running the test, place the HGM-111 meter, test strips and
glucose control solution at room temperature for 30 minutes to allow
them to reach the new temperature before use.
◆
When to run the glucose control solution test
:
□ The test strip vial is opened for long time or you think the test strips
have been damaged.
□ When the meter could be damaged or not working properly.
□ When you want to double check on the meter’s performance.
Chapter 4 Glucose Control Solution Testing (Optional)
Warning: If you will like to perform control solution test or
have any question, please contact your local
authorized distributor for more information.
23
BLOOD GLUCOSE MONITORING SYSTEM-HGM-111